Mako Surgical Corporation: RESTORIS PST RIO Offset Shell Impactor For use by orthopedic surgeons specifically for the purpose of positioning and implanting acetabular cup components during Total Hip Arthroplasty (THA) surgery.

Recall number
Z-1583-2016
Recalling firm
Mako Surgical Corporation
Product
RESTORIS PST RIO Offset Shell Impactor For use by orthopedic surgeons specifically for the purpose of positioning and implanting acetabular cup components during Total Hip Arthroplasty (THA) surgery.
Status
Terminated
Initiated
2016-02-23
Posted
not on file
Terminated
2016-12-14
Reason
Potential disassociation of the orientation pin.
Root cause
Nonconforming Material/Component
Action
On March 4, 2016 Stryker Orthopaedics initiated a voluntary product recall of the RESTORIS PST RIO Offset Shell Impactor informing the customer in an Urgent Product Recall to do the following: 1. Inspect all metal instruments for any damage or difficulty disassembling/assembling instruments 2. Return all affected products at their location to Stryker C/O Stericycle, 2670 Executive Dr. Suite A, Indianapolis, IN, 46241, Attn of Event 8763, Clayton Odor. 3. Complete and sign the Business Reply Form and send it back via e-mail or fax within five business days. Contact 866-241-0741 or via e-mail strykerortho8763@stericycle.com
Quantity
96
Distribution
Nationwide Distribution -- AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, and WI.
Lot, serial or model codes
Lot #'s 28548, 29130, 29479, 30440, 31801, 31802, and 32145.
510(k) numbers
["K121064"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3005985723
Firm city
Plantation
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register