FDA Medical Device Recalls (openFDA)
Mako Surgical Corporation: RESTORIS PST RIO Offset Shell Impactor For use by orthopedic surgeons specifically for the purpose of positioning and implanting acetabular cup components during Total Hip Arthroplasty (THA) surgery.
- Recall number
- Z-1583-2016
- Recalling firm
- Mako Surgical Corporation
- Product
- RESTORIS PST RIO Offset Shell Impactor For use by orthopedic surgeons specifically for the purpose of positioning and implanting acetabular cup components during Total Hip Arthroplasty (THA) surgery.
- Status
- Terminated
- Initiated
- 2016-02-23
- Posted
- not on file
- Terminated
- 2016-12-14
- Reason
- Potential disassociation of the orientation pin.
- Root cause
- Nonconforming Material/Component
- Action
- On March 4, 2016 Stryker Orthopaedics initiated a voluntary product recall of the RESTORIS PST RIO Offset Shell Impactor informing the customer in an Urgent Product Recall to do the following: 1. Inspect all metal instruments for any damage or difficulty disassembling/assembling instruments 2. Return all affected products at their location to Stryker C/O Stericycle, 2670 Executive Dr. Suite A, Indianapolis, IN, 46241, Attn of Event 8763, Clayton Odor. 3. Complete and sign the Business Reply Form and send it back via e-mail or fax within five business days. Contact 866-241-0741 or via e-mail strykerortho8763@stericycle.com
- Quantity
- 96
- Distribution
- Nationwide Distribution -- AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, and WI.
- Lot, serial or model codes
- Lot #'s 28548, 29130, 29479, 30440, 31801, 31802, and 32145.
- 510(k) numbers
- ["K121064"]
- PMA numbers
- not on file
- Product code
- OLO
- Firm FEI
- 3005985723
- Firm city
- Plantation
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28