Philips North America Llc: Veradius Unity. Product Code 718132. The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the devices.

Recall number
Z-1582-2025
Recalling firm
Philips North America Llc
Product
Veradius Unity. Product Code 718132. The devices are used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the devices.
Status
Open, Classified
Initiated
2025-02-25
Posted
2025-04-21
Terminated
not on file
Reason
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
Root cause
Labeling design
Action
Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and maintenance instructions. Philips proposes sending the Electronic Product Radiation Defect notification to customers immediately upon FDA approval of the corrective action plan, and completion of the remaining activities within 4 months of FDA approval of the corrective action plan. This corrective action does not require any customer transportation reimbursement. The following statement is included in the customer communication: The manufacturer will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. Draft of letter provided. Philips commits to notifying the FDA on a monthly basis providing progress with this field correction.
Quantity
not on file
Distribution
US
Lot, serial or model codes
Product UDI: (01)00884838058194(21) 459801442981, Wireless foot switch set 3P (WFSS 3P) 459801238231, Wireless foot switch 3P, 459801257861, WFS Base Station 459800627394, Wireless foot switch set 3P 459800415535, Wireless foot switch 3P, 459800415542, WFS Base Station
510(k) numbers
K142708
PMA numbers
not on file
Product code
OWB
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source