Brainlab AG: ExacTrac 6.x. is software used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures.

Recall number
Z-1582-2015
Recalling firm
Brainlab AG
Product
ExacTrac 6.x. is software used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures.
Status
Terminated
Initiated
2015-03-06
Posted
2015-05-06
Terminated
2017-04-24
Reason
ExacTrac 6.x Patient Positioning System: Potentially incorrect patient positioning when using the ExacTrac Cone Beam CT (CBCT) with a CBCT acquired at a couch angle other than 0.0 degrees.
Root cause
Software in the Use Environment
Action
FIELD SAFETY / PRODUCT NOTIFICATION letters dated March 6, 2015 were sent to all customers of record. The letters included instructions for customers to: 1) With immediate effect, always make sure the couch is at exactly 0.0 degrees (displayed on the Varian monitor) when acquiring a CBCT set to import into the ExacTrac software; 2) For Varian systems providing an option to prevent a CBCT acquisition for couch angles other than 0.0 degrees, Brainlab recommends selecting that option for use in combination with ExacTrac; and, 3) Continue to verify any ExacTrac CBCT based correction using the ExacTrac X-ray verification and/or an external IGRT procedure as mandatory. The letters further advised that Brainlab will be taking the following actions: 1. Brainlab provides existing potentially affected ExacTrac v.6.x CBCT Import & Alignment Software customers (with a Varian C-Series linac) with this product notification information; and, 2. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact the customers starting August 2015 to schedule the update installation. Customers with questions about this issue were advised to contact Brainlab at 800-597-5911 or via e-mail by writing at us.support@brainlab.com.
Quantity
25 systems (US only); 53 systems (Foreign)
Distribution
*** US: Nationwide; *** FOREIGN: Austria, China, Denmark, France, Germany, Hong Kong, Japan, Netherlands, Qatar, Russian Federation, South Korea, Switzerland
Lot, serial or model codes
Software Versions: ExacTrac v. 6.x (v. 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0);   Model/Catalog Numbers: 49971, CBCT IMPORT & ALIGNMENT SOFTWARE; 20836, EXACTRAC 6.0 VARIAN CBCT LICENSE;  48320, ET UPGRADE SOFTWARE 6.0 TO 6.1;  48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA;  49973B, ET DATA PREP / REVIEW SYSTEM;  49973A, ET DATA PREP / REVIEW SYSTEM;  49973, ET DATA PREP / REVIEW SYSTEM;  49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;  48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT;  48302, ET DATA PREP / REVIEW SYSTEM
510(k) numbers
["K120789"]
PMA numbers
not on file
Product code
IYH
Firm FEI
3002619595
Firm city
Feldkirchen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register