Depuy Orthopaedics, Inc.: PolyAx Locking Plate System, POLYAX FEMORAL PLATE - LEFT, 6 hole, Sterile, REF 8141-31-106, Depuy Orthopaedics, Inc., Warsaw, IN.

Recall number
Z-1582-2011
Recalling firm
Depuy Orthopaedics, Inc.
Product
PolyAx Locking Plate System, POLYAX FEMORAL PLATE - LEFT, 6 hole, Sterile, REF 8141-31-106, Depuy Orthopaedics, Inc., Warsaw, IN.
Status
Terminated
Initiated
2010-11-01
Posted
2011-03-07
Terminated
2012-06-27
Reason
The firm determined that the 8.0 mm hole threads in the PolyAx Distal Femoral Locking Plate were incorrectly manufactured, which prevents the mating screw from locking in place. The screw should lock with the plate, instead the screw passes through the plate.
Root cause
Other
Action
Urgent Information - Recall Notice letters, dated November 1, 2010 were sent to customers. The letters described the reason for recall and identified the product affected along with the affected units. The letters also provided clinical guidance and stated that the problem had been corrected and that DePuy was in the process of replenishing its inventory. Hospitals were to determine if any of the recalled components they received were still on hand, and return them immediately to their DePuy Representative or directly to DePuy for credit.
Quantity
246 all products
Distribution
Nationwide Distribution -- ME, MA, FL, VA, CO, TX, WY, AZ, MO, WI, GA, PA, AL, NY, OH, NJ, NC, and CA.
Lot, serial or model codes
DLHCRB.
510(k) numbers
["K060969"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register