Skip to the content

FDA Medical Device Recalls (openFDA)

Resource Optimization & Innovation Llc: Regard, Item Number: 800410003, Sterile, CV0479C - Open Heart PK - Spohn

Recall number
Z-1580-2017
Recalling firm
Resource Optimization & Innovation Llc
Product
Regard, Item Number: 800410003, Sterile, CV0479C - Open Heart PK - Spohn
Status
Terminated
Initiated
2017-01-09
Posted
not on file
Terminated
2018-09-28
Reason
Custom procedure trays contain a component, a light glove, that has been recalled by the manufacturer because it can split during application to a light handle breaking the sterile field.
Root cause
Nonconforming Material/Component
Action
ROi sent an Urgent Recall Notice dated January 9, 2017, to all affected consignees. The recalling firm instructed consignees to quarantine affected product and complete the response form. The recalling firm stated that they will issue over lableing for the affected packs. Sterile replacement light covers will be sent directly to the end user if available. Consignees were instructed to inform their subaccounts of the recall. For questions regarding this recall call 417-820-2793.
Quantity
117 kits
Distribution
US Distribution to LA and TX.
Lot, serial or model codes
028316C Exp Date 2017-03-30; 028909C Exp Date 2017-03-30
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
3004513970
Firm city
Springfield
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register