Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: TumorLoc Software Application, Software versions: 23.1.1, 20.7.13, and 20.8.15. This is an option available for purchase for use on the GEMINI GXL 16 Slice PET/CT System (contains EBW 455011002031 as an accessory), Catalog #882410, Model Number: 4535 679 71891. CT host. 2.2. Tumor Localization is an application used in radiation oncology departments for localizing target volumes for radiation therapy planning. It includes exclusive features for viewing Respiratory Correlated CT data sets and analyzing motion of target and surrounding anatomy.

Recall number
Z-1579-2009
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
TumorLoc Software Application, Software versions: 23.1.1, 20.7.13, and 20.8.15. This is an option available for purchase for use on the GEMINI GXL 16 Slice PET/CT System (contains EBW 455011002031 as an accessory), Catalog #882410, Model Number: 4535 679 71891. CT host. 2.2. Tumor Localization is an application used in radiation oncology departments for localizing target volumes for radiation therapy planning. It includes exclusive features for viewing Respiratory Correlated CT data sets and analyzing motion of target and surrounding anatomy.
Status
Terminated
Initiated
2009-02-19
Posted
2009-07-06
Terminated
2012-09-21
Reason
A software anomaly in the TumorLoc software for the firm's Brilliance CT scanners was discovered. The same software anomaly was subsequently determined to be present in the firm's GEMINI PET/CT units. The software defect results in a the potential for data to be lost during reconstruction of images in which slices can either be discarded, or interleaved based upon their series ID.
Root cause
Software design
Action
An "URGENT - Field Safety Notice" dated February 12, 2009 was sent via FedEx to domestic customers. The notice described the affected products, issue, involved hazard should problem occur and corrective actions for the customer/user. The letter also notified users/customers that a Philips Healthcare Field Service Engineer will contact and visit their site within 6 months to install a TumorLoc upgrade. Notification will also be sent to the international consignees by Philips personnel. For further information or support concerning this issue, please contact your local Philips representative: Customer Care Center (1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: Oncology) or to your local Philips Healthcare office.
Quantity
3 units
Distribution
Worldwide Distribution -- US, Columbia, Germany, France, Netherlands, Canada, Belgium and Slovenia.
Lot, serial or model codes
GEMINI GXL 16 Slice - Serial Numbers: 4149, 4150, and 4173.
510(k) numbers
["K051170"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register