Micro-X Ltd.: Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A

Recall number
Z-1578-2025
Recalling firm
Micro-X Ltd.
Product
Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A
Status
Open, Classified
Initiated
2025-03-28
Posted
2025-04-16
Terminated
not on file
Reason
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
Root cause
Device Design
Action
On March 28, 2025, Micro-X Ltd. issued a "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. Micro-X asked consignees to take the following actions: 1. Continue to use your product as normal in accordance with its intended use. 2. Your local service representative will be in contact to apply this update to the systems at the next site visit. 3. Circulate this notice to all users of this device so they are aware of the issue. 4. If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. 5. Please retain this letter with your system(s) documentation. 6. Please complete and return the attached response form to Micro-X to the Micro-X LTD in Australia: support@micro-x.com and copy djerkovic@micro-x.com.
Quantity
31 units
Distribution
US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US territory).
Lot, serial or model codes
Serial Number/UDI: MSN0334 (01)09357123000051(11)220921(21)00334; MSN0350 (01)09357123000037(11)230328(21)00350 ; MSN0352 (01)09357123000051(11)230823(21)00352; MSN0364 01)09357123000051(11)230705(21)00364; MSN0382 (01)09357123000051(11)230809(21)00382; MSN0383 (01)09357123000051(11)230814(21)00383; MSN0384 (01)09357123000051(11)230816(21)00384; MSN0385 (01)09357123000051(11)230906(21)00385; MSN0386 (01)09357123000051(11)230908(21)00386; MSN0387 (01)09357123000051(11)230914(21)00387; MSN0388 (01)09357123000051(11)230916(21)00388; MSN0389 (01)09357123000051(11)230918(21)00389; MSN0390 (01)09357123000051(11)231013(21)00390; MSN0395 (01)09357123000051(11)231106(21)00395; MSN0397 (01)09357123000051(11)231118(21)00397; MSN0399 (01)09357123000051(11)240122(21)00399; MSN0400 (01)09357123000051(11)240207(21)00400; MSN0401 (01)09357123000051(11)240131(21)00401; MSN0402 01)09357123000051(11)240202(21)00402; MSN0403 (01)09357123000051(11)240205(21)00403; MSN0405 (01)09357123000051(11)240206(21)00405; MSN0408 (01)09357123000051(11)240308(21)00408; MSN0409 (01)09357123000051(11)240312(21)00409; MSN0410 (01)09357123000051(11)240318(21)00410; MSN0411 (01)09357123000051(11)240319(21)00411; MSN0422 (01)09357123000051(11)240724(21)00422; MSN0423 (01)09357123000051(11)240808(21)00423; MSN0424 (01)09357123000051(11)240809(21)00424; MSN0425 (01)09357123000051(11)240812(21)00425; MSN0426 (01)09357123000051(11)240812(21)00426; MSN0427 (01)09357123000051(11)240812(21)00427;
510(k) numbers
["K211423"]
PMA numbers
not on file
Product code
IZL
Firm FEI
3012855798
Firm city
Tonsley
Firm state
not on file
Firm country
Australia

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register