Zimmer, Inc.: NexGen Augment; 5mm Distal Femoral Augment;Part number 00-5990-036-10. Intended to be used in total knee arthroplasty when there is insufficient bone in the distal femur.
- Recall number
- Z-1575-2015
- Recalling firm
- Zimmer, Inc.
- Product
- NexGen Augment; 5mm Distal Femoral Augment;Part number 00-5990-036-10. Intended to be used in total knee arthroplasty when there is insufficient bone in the distal femur.
- Status
- Terminated
- Initiated
- 2015-04-02
- Posted
- 2015-05-01
- Terminated
- 2015-09-23
- Reason
- The affected lot is being recalled because it was packaged without the locking screw. The locking screw packaged with the augment is used to fix the augment to the femoral component. This has the potential to delay surgery
- Root cause
- Mixed-up of materials/components
- Action
- On 4/2/2015, URGENT MEDICAL DEVICE RECALL LOT SPECIFIC notifications were sent to the affected distributors with instructions for locating, quarantining, and returning the affected product. All distributors were notified via electronic mail, and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
- Quantity
- 20 units
- Distribution
- Distributed in the states of IN, MN, FL, and IL, and the countries of Argentina and Germany.
- Lot, serial or model codes
- Part #00-5990-036-10; lot 62815809
- 510(k) numbers
- ["K960279"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28