Skip to the content

FDA Medical Device Recalls (openFDA)

Enpath Medical, Inc.: Enpath Steerable Sheath, model 10775-003, Sterile EO. Single Use Only. Enpath Medical, Inc., 15301 Highway 55 West, Minneapolis, MN 55447, USA. The Enpath Deflectable Catheter is a felexible tipped percutaneous catheter designed for gaining access to the peripheral and coronary systems. The device features adjustable tip geometry through use of a sliding mechanism with locking feature to deflect the catheter. There are two versions of the tool: one with a luer fitting for Tuohy-Borst valve connection, the other with a hemostasis valve and a sideport infusion line. The kit includes a flexible catheter dilator to facilitate deflectable catheter passage.

Recall number
Z-1573-05
Recalling firm
Enpath Medical, Inc.
Product
Enpath Steerable Sheath, model 10775-003, Sterile EO. Single Use Only. Enpath Medical, Inc., 15301 Highway 55 West, Minneapolis, MN 55447, USA. The Enpath Deflectable Catheter is a felexible tipped percutaneous catheter designed for gaining access to the peripheral and coronary systems. The device features adjustable tip geometry through use of a sliding mechanism with locking feature to deflect the catheter. There are two versions of the tool: one with a luer fitting for Tuohy-Borst valve connection, the other with a hemostasis valve and a sideport infusion line. The kit includes a flexible catheter dilator to facilitate deflectable catheter passage.
Status
Terminated
Initiated
2005-08-31
Posted
2005-09-21
Terminated
2006-07-22
Reason
Enpath has become aware that some of the Enpath Medical 8 Fr Steerable sheath devices may have non-conforming hemostasis valves in that the valve can leak or the valve was observed to have a small hole.
Root cause
Other
Action
Clinical locations were sent an Urgent Medical Device Recall letter dated 09/02/05 from Enpath Medical, Inc. The letter states the issue, affected lots to be segregated and to discontinue use of. A Bard Electrophysiology Territory Manager will retrieve affected product.
Quantity
140
Distribution
MA
Lot, serial or model codes
W34968 and W35440
510(k) numbers
["K043489"]
PMA numbers
not on file
Product code
DYB
Firm FEI
2183787
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register