Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular Us Sales, Llc: MAQUET GETINGE GROUP Arterial Cannula OD Used to introduce blood and sterile solutions into the systemic circulation during pediatric extracorporeal circulation procedures.

Recall number
Z-1572-2013
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Product
MAQUET GETINGE GROUP Arterial Cannula OD Used to introduce blood and sterile solutions into the systemic circulation during pediatric extracorporeal circulation procedures.
Status
Terminated
Initiated
2013-05-14
Posted
2013-06-21
Terminated
2014-09-18
Reason
The specified part numbers and lot numbers of Maquet Pediatric Arterial Cannula Packs may contain an incorrectly sized connector.
Root cause
Packaging process control
Action
Maquet sent an "Urgent Product Recall"/"Recall - Device Corrective Action Response" forms dated May 14, 2013, via Federal Express to all affected customers. The notification identifies affected product; informs customers of the nature of the issue; advises of potential adverse patient consequences that may occur; and advises customers to follow instructions on the "Recall-Device Corrective Action Response" forms included as part of the notification. If you have any further questions please contact your local MAQUET representative or call Customer Service at (888) 627-8383.
Quantity
100 units (60 Domestic; 40 International)
Distribution
Worldwide Distribution - US Distribution including the states of Delaware, Illinois, Minnesota , New York., Ohio and Pennsylvania and the countries of Germany, Turkey and France.
Lot, serial or model codes
Catalogue Number:Part Number: Lot Number are as follows: (1) 161410 (10 FR.):701022682:92056588 (2) 163610 (10 Fr):70102597:92059789 (3) 163608 (8 Fr.):70102.2681:92065640 (4) 163606 (6 Fr.):70102.2680:92063566
510(k) numbers
not on file
PMA numbers
not on file
Product code
FGY
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register