Skip to the content

FDA Medical Device Recalls (openFDA)

Becton, Dickinson and Company, BD Biosciences: BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.

Recall number
Z-1570-2016
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
Status
Terminated
Initiated
2016-03-31
Posted
not on file
Terminated
2017-05-26
Reason
One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.
Root cause
Packaging
Action
BD Biosciences sent an Urgent Product Recall letter dated March 2016 to all affected customers on March 31, 2016 by Certified Mail, e-mail and direct delivery by Sales Associates. Letters instructed that any remaining inventory of the affected lot be discarded and that date files obtained using reagents from the affected lot be reviewed for the possibility of incorrect data. Customers with questions were instructed to contact BD Customer Support at 877-232-8995 (prompt3, 3) in the united State. For customers outside the US, customers should contact their local BD Biosciences representative or distributor. For questions regarding this recall call 408-954-6080.
Quantity
1743; 467 US and 1276 International
Distribution
Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.
Lot, serial or model codes
Model: 340334; Manufacaturing lot: 534964
510(k) numbers
["K970836"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2916837
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register