FDA Medical Device Recalls (openFDA)
Luxottica Optical Manufacturing: Polycarbonate prescription single vision lenses for use in sport goggles. To be worn by a customer in a sport goggle frame to provide corrective refraction in accordance with the customer's prescription. Per the customer's request, some of the devices may also provide protection from bright sunlight.
- Recall number
- Z-1568-2011
- Recalling firm
- Luxottica Optical Manufacturing
- Product
- Polycarbonate prescription single vision lenses for use in sport goggles. To be worn by a customer in a sport goggle frame to provide corrective refraction in accordance with the customer's prescription. Per the customer's request, some of the devices may also provide protection from bright sunlight.
- Status
- Terminated
- Initiated
- 2010-06-30
- Posted
- 2011-03-07
- Terminated
- 2012-09-19
- Reason
- The polycarbonate prescription lenses may not meet the frame manufacturer's minimum lens thickness specification of 2.0 millimeters. LNAD made the decision to remove these sport goggles from the market because of the potential risk that the lenses could break or disengage from the frame in the event of an impact.
- Root cause
- Other
- Action
- Luxottica Optical Manufacturing sent notification letters on June 30, 2010, to their customers informing them of the problem. Customers were instructed to complete the enclosed Sport Goggle Return Form, and to indicate whether they would like a replacement pair of goggles or a refund. Place the completed Sport Goggle Return Form and their sport goggles (without the sport goggle case) in the envelope provided and return it to the firm via the United States Postal Service. Customers were instructed not to return to the retail store where purchased. If customers were unable to return their sport goggles because they were lost or discarded, customers should complete the appropriate information on the Sport Goggle Return Form, being sure to indicate whether they would like a replacement pair of goggles or a refund, and then sign the form. Place the completed Sport Goggle Return Form in the envelope provided and return it to the firm via the United States Postal Service. For any questions, call 1-877-488-2407. They instructed their customers to return the product for replacement or refund.
- Quantity
- 2,916 individual lenses (1,458 pair of sport goggles)
- Distribution
- Worldwide Distribution - USA including the following states: AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI & WV; and the country of Canada.
- Lot, serial or model codes
- In the U.S., the Liberty Sport frame models and the corresponding SKU numbers into which the lenses at issue were inserted are: MaxMorpheus - 3703097, 3703105, 3703121, 3703139, 3703154, 3703162; Slam - 4286102, 4286103; Rec Spec Maxx - 1292931, 2102549, 2102549, 2102556, 2102564, 2102606, 2621068, 2621076, 2621084, 3300266, 3300274, 3300282, 3300332, 3300340, 3300357; Rec Spec Zone - 2102622, 3300407; Rec Spec Fusion - 2102655, 2102663, 3300373; Rec Spec Matrix - 1270439, 1278090, 1278089; Rec Spec HS2C - 1260789; Rec Spec 2003 - 1278092; Rec Spec Soft Touch - 1260791 In Canada, the Liberty Sport frame models and the corresponding SKU numbers into which the lenses at issue were inserted are: MaxMorpheus - 3703097, 3703162; Rec Spec Maxx - 2102549, 2102556, 2621068, 2621076, 3300381.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HQG
- Firm FEI
- 3008153912
- Firm city
- Mason
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28