LeMaitre Vascular, Inc.: LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the removal of emboli and thrombi during embolectomy and/or thrombectomy. It can also be used for catheter placement over a guidewire, vessel occlusion, fluid infusion and/or aspiration.

Recall number
Z-1567-2020
Recalling firm
LeMaitre Vascular, Inc.
Product
LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the removal of emboli and thrombi during embolectomy and/or thrombectomy. It can also be used for catheter placement over a guidewire, vessel occlusion, fluid infusion and/or aspiration.
Status
Open, Classified
Initiated
2020-03-23
Posted
2020-04-01
Terminated
not on file
Reason
There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.
Root cause
Process control
Action
On March 23, 2020, the firm distributed "URGENT: MEDICAL DEVICE RECALL" letters to affected customers. Customers were notified of the risk of the catheter failing to deflate, which may ultimately lead to tip separation. Customers were asked to do the following: 1. Check your inventory and immediately quarantine any recalled devices. 2. Complete the form at the end of this letter. Please note that you must return the form even if you have no devices in inventory. 3. Scan the reply form and send it to recalls@lemaitre.com. 4. If you have recalled devices at your facility, LeMaitre Vascular, Inc. will send you instructions for returning the devices. 5. When the recalled devices have been received by LeMaitre Vascular, we will arrange for replacement material. 6. If you have transferred devices to another facility, please forward a copy of this recall letter to them. Customers with questions may contact the Senior Quality Engineer, at rlerer@lemaitre.com or 781-221-2266. ***Updated 8/21/20*** Firm added Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246 to the recall. These lots were previously on hold in a warehouse in Germany, but in July they were mistakenly distributed to customers in Europe.
Quantity
43315
Distribution
Worldwide distribution: US (Nationwide distribution) and foreign countries of: Australia, Austria, Belgium, Bulgaria, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Georgia, Hungary, Ireland, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, and Vietnam.
Lot, serial or model codes
OTW2897¿ OTW3273¿ OTW3333¿ OTW3379¿ OTW3418¿ OTW3543¿ OTW3564¿ OTW3620¿ OTW3700¿ OTW3740¿ OTW3758¿ OTW3805¿ OTW3865¿ OTW3879¿ OTW3892¿ OTW3966¿ OTW4009¿ OTW4010¿ OTW4017¿ OTW4018¿ OTW4024¿ OTW4060¿ OTW4063¿ OTW4064¿ OTW4065¿ OTW4117¿ OTW4201¿ OTW4238¿ OTW4239¿ OTW2957¿ OTW3001¿ OTW3266¿ OTW3395¿ OTW3565¿ OTW3637¿ OTW3638¿ OTW3658¿ OTW3665¿ OTW3670¿ OTW3671¿ OTW3682¿ OTW3683¿ OTW3694¿ OTW3695¿ OTW3701¿ OTW3708¿ OTW3727¿ OTW3728¿ OTW3747¿ OTW3769¿ OTW3770¿ OTW3775¿ OTW3776¿ OTW3779¿ OTW3780¿ OTW3810¿ OTW3820¿ OTW3821¿ OTW3853¿ OTW3854¿ OTW3880¿ OTW3881¿ OTW3886¿ OTW3893¿ OTW3898¿ OTW3899¿ OTW3915¿ OTW3916¿ OTW3924¿ OTW3930¿ OTW3957¿ OTW3958¿ OTW3981¿ OTW3982¿ OTW4011¿ OTW4025¿ OTW4026¿ OTW4030¿ OTW4031¿ OTW4052¿ OTW4053¿ OTW4054¿ OTW4066¿ OTW4067¿ OTW4068¿ OTW4073¿ OTW4074¿ OTW4075¿ OTW4092¿ OTW4093¿ OTW4098¿ OTW4099¿ OTW4105¿ OTW4106¿ OTW4107¿ OTW4119¿ OTW4120¿ OTW4157¿ OTW4158¿ OTW4188¿ OTW4197¿ OTW4220¿ OTW4221¿ OTW4222¿ OTW4229¿ OTW4230¿ OTW4247¿ OTW4249¿ OTW4254¿ OTW4255¿ OTW4258¿ OTW4259¿ OTW4260¿ OTW4261¿ OTW4263¿ OTW4291¿ OTW4294¿ OTW4339¿ OTW4341¿ OTW4363¿ OTW4364¿ OTW2942¿ OTW2959¿ OTW2987¿ OTW3207¿ OTW3292¿ OTW3303¿ OTW3312¿ OTW3322¿ OTW3323¿ OTW3351¿ OTW3367¿ OTW3438A¿ OTW3511¿ OTW3635¿ OTW3636¿ OTW3660¿ OTW3661¿ OTW3667¿ OTW3688¿ OTW3689¿ OTW3706¿ OTW3707¿ OTW3711¿ OTW3717¿ OTW3718¿ OTW3734¿ OTW3750¿ OTW3783¿ OTW3784¿ OTW3794¿ OTW3795¿ OTW3827¿ OTW3828¿ OTW3851¿ OTW3852¿ OTW3862¿ OTW3863¿ OTW3882¿ OTW3896¿ OTW3897¿ OTW3907¿ OTW3908¿ OTW3931¿ OTW3936¿ OTW3944¿ OTW3945¿ OTW3962¿ OTW3963¿ OTW3978¿ OTW3979¿ OTW3995¿ OTW3996¿ OTW4033¿ OTW4034¿ OTW4057¿ OTW4058¿ OTW4084¿ OTW4085¿ OTW4127¿ OTW4128¿ OTW4140¿ OTW4141¿ OTW4146¿ OTW4147¿ OTW4176¿ OTW4177¿ OTW4183¿ OTW4184¿ OTW4191¿ OTW4192¿ OTW4193¿ OTW4232¿ OTW4317¿ OTW4336¿ OTW2872¿ OTW3191¿ OTW3439¿ OTW3537¿ OTW3623¿ OTW3662¿ OTW3668¿ OTW3669¿ OTW3690¿ OTW3691¿ OTW3712¿ OTW3735¿ OTW3766¿ OTW3771¿ OTW3812¿ OTW3822¿ OTW3823¿ OTW3836¿ OTW3849¿ OTW3850¿ OTW3874¿ OTW3888¿ OTW3903¿ OTW3925¿ OTW3932¿ OTW3946¿ OTW3947¿ OTW3968¿ OTW3969¿ OTW3987¿ OTW3988¿ OTW4020¿ OTW4021¿ OTW4036¿ OTW4037¿ OTW4038¿ OTW4079¿ OTW4080¿ OTW4087¿ OTW4196¿ OTW4267¿ OTW4268¿ OTW4269¿ OTW4270¿ OTW4275¿ OTW4276¿ OTW4277¿ OTW4288¿  ***Added 3/19/20*** OTW2847¿ OTW2863¿ OTW2871¿  ***Updated 4/14/20*** The firm noted upon review that some of the original recall lots were originally relabeled with different lot numbers for Non-US customers. Those additional lot numbers are as follow:  OTW3461 OTW3501 OTW3591 OTW3628 OTW3737 OTW3742 OTW3753 OTW3813 OTW3859 OTW3861 OTW4208  ***Updated 8/21/20***  Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246
510(k) numbers
["K022145"]
PMA numbers
not on file
Product code
DXE
Firm FEI
1220948
Firm city
Burlington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register