GE Healthcare: GE Healthcare Precision RXi version e Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, and 50KW or 65 kW pulsing generator. The system is configured with a 23 cm image intensifier and a 512x512 Image Chain.

Recall number
Z-1565-2008
Recalling firm
GE Healthcare
Product
GE Healthcare Precision RXi version e Remote X-Ray Imaging R&F System. It is a remote analog R&F system consisting of a tilting table with integrated spot film device, and 50KW or 65 kW pulsing generator. The system is configured with a 23 cm image intensifier and a 512x512 Image Chain.
Status
Terminated
Initiated
2007-12-18
Posted
2008-08-24
Terminated
2012-08-09
Reason
GE Healthcare Precision RXi tilt joystick became stuck in the Trendelenburg direction causing the patient to slide off the table. Also recall conducted to investigate the possibility of accidentally activation of the remote console joystick where a fatal accident occurred on a non-GE Remote R&F System. .
Root cause
Device Design
Action
Two letters "GE Product Safety Notification" dated 12/18/07 and 3/7/08, addressed to Hospital Administrators, Manager of Radiology and Radiologists, were sent to each consignee. Each letter addressed a separate safety issue. Letters described the affected product, safety instructions, corrective action and contact information. The joystick can become stuck causing unintentional motion of the tube. Field Modification Instructions (FMI 10802/10803) have been issued to the field to install additional warning labels on the patient table, install a protector for the console joystick and inspect and correct the clearance between the console joystick and its mounting location. If your facility has a Precision RXi or Precision e system manufactured before March 2007, a GE Healthcare Field Engineer will schedule a field visit to install the added joystick protection. This upgrade will be performed at no charge. If you have any questions/concerns regarding this issue, please contact your local GE Healthcare sales or service representative.
Quantity
6
Distribution
(USA) AK CO IL LA MA MI MS NC NJ NY PA SC TN UT VA WA WI (OUSA) BR CA MX AU C3 KR PH TW AE AT BE BF BG CM CY CZ DE ES FI FR GB GE GR GY HR HU IT LB LV MT NA NG PL PT RO RU SA SD SE SI SK SY TN TR UA YU ZA
Lot, serial or model codes
10805 10805 10805 10804 10804 10804
510(k) numbers
["K041605"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register