Skip to the content

FDA Medical Device Recalls (openFDA)

Straumann USA LLC: TLX/TLC SP Guided Implant Driver, for ratchet, stainless steel; REF: 037.3002;

Recall number
Z-1561-2025
Recalling firm
Straumann USA LLC
Product
TLX/TLC SP Guided Implant Driver, for ratchet, stainless steel; REF: 037.3002;
Status
Open, Classified
Initiated
2025-03-08
Posted
2025-04-14
Terminated
not on file
Reason
The devices are missing the laser marked depth markings.
Root cause
Process control
Action
On March 8, 2024 URGENT FIELD SAFETY ACTION letters were sent to customers. Action to be taken: 1. If Article 037.3002 from Lots LGMZ8 and GLKJ9 is still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If Article 037.3002 from Lots LGMZ8 and GLKJ9 is still in your inventory, return it for a replacement as indicated on the Customer Confirmation Form 3. In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC 60 Minuteman Road Andover, MA 01810 Transmission of the Field Safety Notice This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action We apologize for any inconvenience that this may cause.
Quantity
45 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, IN, ME, MN and the countries of Belgium, Finland, France, Germany, Iceland, Israel, Italy, Latvia, Malta, Romania, Switzerland, Ukraine, United Kingdom.
Lot, serial or model codes
REF: 037.3002; UDI-DI: 07630031750587; Packaging Lot number: LGMZ8, GLKJ9; Lot number engraved on part: HTXT5.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDP
Firm FEI
1000121052
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register