FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.
- Recall number
- Z-1561-2016
- Recalling firm
- Elekta, Inc.
- Product
- Radionuclide Radiation Therapy System Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.
- Status
- Terminated
- Initiated
- 2016-04-19
- Posted
- not on file
- Terminated
- 2021-07-16
- Reason
- The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.
- Root cause
- Use error
- Action
- Elekta sent an Important User Notice (100-03-202-27) letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the Acknowledgement Form attached. For questions contact your local Elekta office.
- Quantity
- 217
- Distribution
- Worldwide Distribution - US Nationwide in the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WV, WI, and countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom
- Lot, serial or model codes
- Perfexion - 14753-099/6052, 12579-007/6250, 10394-LGK01/6144, 11920-C5210/6088, 17397-001/6158, 105528-001/6122, 14857-001/6028, 12236-006/6097, 12473-003/6005, 14857-001/6028, 16867-001/6166, 12668-003/6145 10208-003/6215, 10698-001/6237, 10979-002/6213, 10316-500/6210, 21061-001/6195, 20364-001/6103, 11720-002/6143, 12620-002/6120, 10611-001/6081, 12846-004/6064, 11524-002/8001, 11420-001/6212, 11420-001/6228, 14983-001/6076, 11083-HOM001/6070, 10247-002/6136,10684-002/6090,11868-004/6002, 10795-K08002/6063, 20354-ENS/6160, 11792-005/6077, 11291-001/6018, 10326-003/6105, 18630-001/6031, 17322-001/6121, 30000656-002/6227, 12629-003/6026, 10924-001/6089, 11154-003/6050, 20438-003/6013 17522-001/6214, 17525-002/6038, 11414-LGK002/6205, 11940-LGK002/6181, 10514-LGK002/6183, 11639-LGK001/6042, 11390-LGK001/6109, 10416-LGK001/6125, 11670-LGK001/6133, 11990-LGK001/6118, 10648-LGK001/6098, 15104-LGK001/6080, 10654-LGK001/6096, 10959-LGK001/6132, 10997-LGK001/6087, 15748-LGK001/6106, 11149-LGK002/6206, 11986-LGK002/6218, 10929-LGK002/6175, 10879-LGK002/6234, 11434-LGK002/6255, 11991-LGK002/6198, 13480-LGK002/6148, 10425-LGK002/6254, 10583-LGK002/6161, 11563-LGK002/6162, 10103-LGK02/6149, 10541-LGK002/6142, 10848-LGK001/6108, 11433-LGK001/6116, 14900-LGK001/6058, 15045-LGK001/6091 11784-LGK001/6055, 30000732-001/6223, 40017-002/6170, 12085-006/6163, 12109-002/6169, 12959-002/6033, 10209-001/6046, 10960-005/6193, 0002718-001/6113, 30001751-WOJ701/6068, 12451-001/6019, 14878-003/6067, 105618-001/6110, 19634-001/6155, 16647-001/6187, 14891-001/6104, 30001527-001/6231, 12028-LGK002/6217, 11586-001/6157, 10728-001/6062, 11999-001/6126, 15416-001/6086, 10459-001/6123, 10158-001/6177, 10234-002/6111, 40023-001/6178, 10444-001/6085, 17033-003/6043, 11530-001/6101, 11553-001/6024, 10779-07/6069, 13465-CHUV03/6094, 10424-001/6182, 13370-002/6075, 12214-003/6007, 30000655-001/6199, 10152-001/6159, 11550-001/6065, 10118-001/6230, 12766-001/6112, 13160-001/6039, 11042-001/6003, 12270-001/6017, 18062-001/6146, 11710-004/6141, 11820-006/8006, 11570-004/6220, 11127-009/6219, 12164-004/6221, 10434-004/6244, 10542-012/6238, 13826-002/6239, 10567-041/6191, 98300-003/6140, 10567-037/6119, 14569-001/6135, 10643-005/6117, 11798-009/6060, 17357-001/6130, 11813-003/6034, 10356-007/6079, 10567-045/6232, 10567-032/6074, 10567-030/6036, 10567-025/6029, 11293-004/6004, 12269-002/6082, 11772-004/6048, 10164-004/6009, 10349-003/6066, 10567-026/6011, 10560-003/6127, 11073-008/6020, 10567-036/6102, 14236-001/6114, 10567-035/6092, 10379-002/6061, 11030-006/6030, 13050-001/6037, 10909-009/6010, 11714-003/6072, 12985-003/6025, 14030-002/6015, 10567-031/6054, 10064-008/6021, 13558-001/6099, 10567-038/6107, 11939-003/6008, 10591-002/6083, 10567-024/6014, 11848-003/6095, 11624-003/6078, 10383-007/6100, 11395-010/6032, 11198-005/6124, 10982-003/6053, 11256-001/6012, 10905-004/6016, 10387-011/6093, 11138-004/6115, 10524-003/6073, 10908-003/6045, 10762-005/6057, 10193-007/6056, 10673-016/6006, 11007-005/6035, 11791-006/6022, 11377-003/6186, 12752-004/6242, 11039-006/6188, 10597-002/6167, 17539-001/6185, 14153-001/6197, 30000118-001/6240, 10567-039/6180, 15053-001/6179, 11731-005/6200, 11090-005/6165, 11077-007/6207, 30000171-002/6172, 10760-006/6173, 11733-001/6192, 10817-009/6204, 13514-001/6211, 10877-GK01/6151, 10436-009/6150, 11396-003/6176, 10917-004/6128, 11373-002/6184, 10567-042/6129, 10567-043/6203, 12614-003/6139, 10297-006/6171, 11347-003/6134 LEKSELL GAMMA KNIFE ICON - 11868-008/6131, 11290-ENS1/6174, 12027-EZT-G1/6041, 10444-LGK004/8003, 11352-001/6044, 10622-005/6194, 10886-013/6027, 11201-004/6047, 11834-003/6222
- 510(k) numbers
- ["K151159","K151561"]
- PMA numbers
- not on file
- Product code
- IWB
- Firm FEI
- 1037831
- Firm city
- Atlanta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28