Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Access Systems: PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is intended for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. The PowerLoc MAX SWIS is also indicated for power injection of contrast media into the central venous system only with an implanted port that is also indicated for power injection.

Recall number
Z-1560-2015
Recalling firm
Bard Access Systems
Product
PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is intended for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. The PowerLoc MAX SWIS is also indicated for power injection of contrast media into the central venous system only with an implanted port that is also indicated for power injection.
Status
Terminated
Initiated
2015-03-13
Posted
2015-04-30
Terminated
2015-06-24
Reason
Bard Access Systems is conducting a field action due to the potential that an incorrect Maximum Flow rate label associated to the power injection IFU for the PowerLoc MAX Safety Infusion Set (SWIS) was attached to the outside of the unit pouch.
Root cause
Labeling False and Misleading
Action
The firm, Bard, sent an "URGENT: Notification-Potential Incorrect Infusion IFU Label for PowerLoc MAX Safety Winged Infusion Set (SWIS)" letter dated March 25, 2015 to its customers. The letter described the product, problem and actions taken. The firm has corrected the cause of the issue and all future lots will be manufactured with the correct maximum flow rate of 2mL/second. Customers may return for a replacement or keep items. Please direct any return requests and questions to our coordinator at 801-522-5640.
Quantity
700 units
Distribution
US Distribution to states of: GA and MN.
Lot, serial or model codes
Lot D430107
510(k) numbers
["K073050"]
PMA numbers
not on file
Product code
FPA
Firm FEI
3006260740
Firm city
Salt Lake City
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register