Teleflex Medical: Deknatel Teleflex Medical Stainless Steel 2 (metric 5) Monofilament Stainless Steel Suture, Nonabsorbable Surgical Suture, U.S.P. Used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Recall number
Z-1560-2014
Recalling firm
Teleflex Medical
Product
Deknatel Teleflex Medical Stainless Steel 2 (metric 5) Monofilament Stainless Steel Suture, Nonabsorbable Surgical Suture, U.S.P. Used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Status
Terminated
Initiated
2014-03-11
Posted
2014-05-07
Terminated
2017-08-14
Reason
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
Root cause
Employee error
Action
Urgent Medical Device Recall Notification letters were issued on March 11, 2014 to consignees. The letter requested that customers discontinue use and quarantine any product on hand. The attached Recall Acknowledgment Form should be completed and returned to Teleflex regardless of whether affected product is on hand or not. Once received a customer service representative will contact customers with instructions on how to return the affected product. Questions should be directed to Customer Service at 1-866-246-6990.
Quantity
32,271 ea.
Distribution
Worldwide Distribution -- USA, including the states of AL, CA, FL, GA, IL, IN, KS, LA, ME, MA, MI, MN, MS, MO, NV, NJ, NM, NY, OH, PA, RI, SC, TN, TX, UT, VA, WV, and WI; and countries of Canada, Germany, and Ireland.
Lot, serial or model codes
Product Code X-4981M4 - Lot 02J0800451
510(k) numbers
["K930738"]
PMA numbers
not on file
Product code
GAQ
Firm FEI
3005747797
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register