Medtronic Navigation, Inc: Polaris Spectra Position Sensor Unit, a component of the StealthStation S7 System, Medtronic Navigation, Louisville, CO 80027. The Polaris Spectra Position Sensor Unit (PSU) is the component of the StealthStation S7 System which detects optical markers on instruments and reference frames, determines their spatial positions, and continuously reports this information to the system computer. The computer uses this spatial information, in conjunction with information regarding the geometry of the instrument currently in use, to determine where the tip of the instrument is located on the patient anatomy.
- Recall number
- Z-1558-2011
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Polaris Spectra Position Sensor Unit, a component of the StealthStation S7 System, Medtronic Navigation, Louisville, CO 80027. The Polaris Spectra Position Sensor Unit (PSU) is the component of the StealthStation S7 System which detects optical markers on instruments and reference frames, determines their spatial positions, and continuously reports this information to the system computer. The computer uses this spatial information, in conjunction with information regarding the geometry of the instrument currently in use, to determine where the tip of the instrument is located on the patient anatomy.
- Status
- Terminated
- Initiated
- 2010-12-22
- Posted
- 2011-03-04
- Terminated
- 2011-03-11
- Reason
- Position Sensor Unit may generate tracking errors or may stop tracking immediately.
- Root cause
- Component design/selection
- Action
- The firm, Medtronic, sent a "Product Correction Notification" letter dated December 22, 2010, to all US Consignees/Customers on December 23, 2010 and foreign Consignees/Customers on December 27, 2010. The letter described the product, problem and actions to be taken. The customer were instructed to discontinue use of the impacted navigation system(s) until a replacement camera (PSU) is installed. A Medtronic Navigation Representative will visit each facility to remove the navigation camera and install a replacement. If you have any questions, contact your Medtronic Navigation Representative or Medtronic Navigation Technical Support services at 1-800-595-9709 or +1-720-890-3200.
- Quantity
- 11 (US), 8 (Intnl)
- Distribution
- Worldwide distribution: USA including states of: CA, CO, GA, SC, TN, and WI; and countries including: Australia, Canada, Chile, Russia, Saudi Arabia, and South Africa.
- Lot, serial or model codes
- Serial Numbers: P703048, P703049, P703050, P703054, P703055, P703060, P703061, P703063, P703066, P703068, P703070, P703076, P703077, P703079, P703114, P703140, P703141, P703143.
- 510(k) numbers
- ["K050438"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1000517638
- Firm city
- Louisville
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28