Stryker Endoscopy: 24mm Rotating C-Mount Coupler, Model Number 1088-020-121S2. The device in indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope or arthroscope is indicated for use.

Recall number
Z-1558-2008
Recalling firm
Stryker Endoscopy
Product
24mm Rotating C-Mount Coupler, Model Number 1088-020-121S2. The device in indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope, endoscope or arthroscope is indicated for use.
Status
Terminated
Initiated
2008-03-31
Posted
2008-08-24
Terminated
2008-09-17
Reason
The coupler on this device was manufactured without set-screws that are used to hold and center the rear assembly and the endobody. The coupler is part of the camera system and affects the ability to focus.
Root cause
Nonconforming Material/Component
Action
Only one customer had all of the defective couplers. The firm called this customer on March 27, 2008, and asked that the 28 couplers remaining in their possession be returned.
Quantity
28 units
Distribution
Product was delivered to one consignee in Iowa.
Lot, serial or model codes
Model Number: 1088-020-121S2 Lot Numbers:  Shipped: 07L046134, 07L046144, 07L046154, 07L046164, 07L046184, 07L046194, 07L046214, 07L046224, 07L046234, 07L046244, 07L046254, 07L046274, 07L046284, 07L046294, 07L046304, 07L046324, 07L046334, 07L046344, 07L046364, 07L046374, 07L046384, 07L046394, 07L046404, 07L046414, 07L046424, 07L046454, 07L046464, and 07L046474.
510(k) numbers
["K983566"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register