Abbott Medical: Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The Aveir leadless pacemaker system supports the implantation and use of an Aveir Leadless Pacemaker (LP) within the targeted chamber(s) of the heart for monitoring a patient s heart rate and providing rate responsive bradycardia pacing therapy to regulate heart rate.

Recall number
Z-1557-2025
Recalling firm
Abbott Medical
Product
Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The Aveir leadless pacemaker system supports the implantation and use of an Aveir Leadless Pacemaker (LP) within the targeted chamber(s) of the heart for monitoring a patient s heart rate and providing rate responsive bradycardia pacing therapy to regulate heart rate.
Status
Open, Classified
Initiated
2025-02-15
Posted
2025-04-14
Terminated
not on file
Reason
Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.
Root cause
Process control
Action
On 02/15/2025, the firm communicated with the implanting/following physicians via phone call and in-person meeting. Customers were informed that a subset of Aveir Leadless pacemaker devices manufactured in January 2025 had experienced a departure from manufacturing baseline parameters due to incorrect equipment set-up. As a result of this issue, the glass insulator separating the battery from the electronics may have microcracks that could lead to the reduction in longevity of the device and loss of functionality. Customers are instructed to: Recognizing that each patient requires individual consideration, Abbott recommends: "Consideration of device retrieval and/or replacement (with a leadless or transvenous device), particularly if your patient is pacemaker dependent. "Follow-up actions should consider evaluation of individual patient needs, disease states, and co-morbidities. "Timing of device replacement should depend on the patient s clinical characteristics. Customer were also instructed to acknowledge the customer notification by returning Acknowledgement Form to crmcommercialcompliance@abbott.com Additional notification of customers were on 02/15/2025 and 02/27/2025 via email.
Quantity
6 devices
Distribution
Worldwide - U.S. Nationwide distribution in the states of FL, SC, and WV and the country of Spain. There were five (5) LSP202V devices distributed within the United States (3 in FL, 1 in SC, 1 in WV) that are impacted by this issue, two (2) of which were implanted, see Attachment A.3 for list of serial numbers. There is 1 LSP112V device that is impacted by this issue, but this was distributed outside of the United States. See Attachment A.4 for list of OUS serial numbers.
Lot, serial or model codes
Model Number: LSP202V UDI-DI code: 05415067040725 Serial Numbers: 1357831, 1357869, 1357885, 1357894, 1357895 Model Number: LSP112V UDI-DI code: 05415067034472 Serial Number: 1357956
510(k) numbers
not on file
PMA numbers
["P150035S003"]
Product code
PNJ
Firm FEI
2017865
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register