FDA Medical Device Recalls (openFDA)
Cordis Corporation: CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease.
- Recall number
- Z-1556-2008
- Recalling firm
- Cordis Corporation
- Product
- CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atherosclerotic disease.
- Status
- Terminated
- Initiated
- 2008-03-10
- Posted
- 2008-07-31
- Terminated
- 2009-10-01
- Reason
- Data transcription error. The 10 affected lots of Catalog Number C10040SL were manufactured with a prior version of the guidewire lumen material, which is not validated with the current manufacturing process. This condition may result in a separation of the device. If a separation was to occur, and the separated fragment of lumen material was not retrieved on the guidewire, blockage of the vessel
- Root cause
- Mixed-up of materials/components
- Action
- A letter (RE: Medical Device Recall) dated March 11, 2008, and Acknowledgement Form was sent overnight to the following 6 individuals in each account: Dir. Cardiac Cath Lab, Dir. Radiology Services, Dir. Vascular Services, Risk Manager, Materials Dir., Hospital Administrator; requesting that the affected product be set aside, and that the form be reviewed, signed, and faxed back to Cordis. Affected product is to be either returned or the local sales rep should the contacted to facilitate the return of the affected product. All product will be replaced. Customer Service can be contacted if there are additional questions relative to product return, replacement, or availability.
- Quantity
- 398 units
- Distribution
- Worldwide Distribution --- including USA and countries of Japan and Canada.
- Lot, serial or model codes
- 10 lots. Lot #s: 13326414*, 13326415*, 13327582, 13329185*, 13332369, 13333077*, 13339105*, 13339106*, 13341244, 13343464. (*These six lots have U.S. distribution.)
- 510(k) numbers
- not on file
- PMA numbers
- ["P020036"]
- Product code
- NIO
- Firm FEI
- 1016427
- Firm city
- Miami Lakes
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28