Elekta, Inc.: Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-1555-2011
- Recalling firm
- Elekta, Inc.
- Product
- Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Status
- Terminated
- Initiated
- 2007-08-23
- Posted
- 2011-03-04
- Terminated
- 2011-04-07
- Reason
- Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent within the iCom tolerance table.
- Root cause
- Other
- Action
- The firm, Elekta, sent an "Important Notice A302" dated August 23, 2007 titled "Update to iCom Tables for Desktop Pro when used with Third Party R&V Systems",to all affected customers. On September 28, 2007, Elekta issued a second Important Notice A304 which reiterated the same information as was in Important Notice A302. The notices described the product, problem and actions to be taken. This notice provides a workaround on how to avoid the issue. In order to avoid this issues documented, Elekta recommended that the customers set the tolerance for collimators in the iCom tolerance table to be 3.0 degrees-no less than 3.0 and to use the Linac Record, if configured, as a means of checking the final collimator position. Elekta noted that the issue will be resolved in a future release of Desktop Pro. If you any questions related to this issue, please contact your local Elekta, third party R&V system representative or Elekta Limited at +44 (0)1293 654200.
- Quantity
- 229 units
- Distribution
- Worldwide distribution: USA including Puerto Rico and countries including: Canada and Mexico.
- Lot, serial or model codes
- 134615, 134621, 134091, 136364, 134172, 136285, 134605, 134183, 136167, 136234, 134259, 134115, 134081, 136047, 136063, 134701, 134328, 134391, 134432, 134436, 134438, 136119, 134535, 134561, 134475, 134564, 134154, 134269, 134348, 134399, 136084, 136087, 136100, 134200, 134118, 134822, 134018, 136368, 134241, 134606, 134711, 134473, 134787, 134768, 134571, 134237, 134104, 136009, 134331, 134272, 136322, 136175, 136200, 134503, 136257, 134322, 134781, 134380, 134709, 136108, 134250, 134254, 134258, 134206, 134195, 134094, 134652, 134704, 136107, 134469, 134472, 134483, 134519, 134729, 136109, 134236, 134239, 134197, 134456, 134047, 134092, 134211, 136373, 134275, 134679, 136106, 136163, 134667, 136181, 134169, 134329, 134126, 134090, 134358, 134427, 134626, 134819, 134027, 134265, 134597, 136319, 134220, 134256, 134396, 134804, 134235, 136338, 134137, 134234, 134249, 136034, 136065, 136110, 136267, 136291, 134213, 134219, 136330, 134750, 136097, 136270, 134457, 134078, 134129, 136195, 134375, 134603, 136074, 134075, 134565, 134196, 134428, 134430, 134008, 134382, 134070, 134157, 134772, 136080, 134860, 134401, 134368, 134388, 134631, 134318, 136015, 136096, 134563, 136165, 136240, 134441, 134017, 134048, 134189, 134395, 134324, 134595, 134140, 134212, 136086, 134354, 136178, 136317, 136318, 134445, 136012, 136332, 134321, 136071, 136089, 136292, 136293, 134673, 134685, 136059, 136169, 134073, 134748, 134754, 134378, 134383, 134458, 134192, 134499, 136069, 136092, 134642, 134295, 134671, 136046, 136147, 134062, 136049, 136036, 136152, 136157, 136083, 136081, 134770, 136118, 136156, 136172, 134352, 134548, 134533, 136153, 136269, 136197, 136252, 136286, 136315, 136241, 134811, 134812, 136258, 136312, 134069, 136310, 134864, 134040, 134046, 134307
- 510(k) numbers
- ["K982713"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28