Elekta, Inc.: Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Recall number
Z-1555-2011
Recalling firm
Elekta, Inc.
Product
Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Status
Terminated
Initiated
2007-08-23
Posted
2011-03-04
Terminated
2011-04-07
Reason
Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent within the iCom tolerance table.
Root cause
Other
Action
The firm, Elekta, sent an "Important Notice A302" dated August 23, 2007 titled "Update to iCom Tables for Desktop Pro when used with Third Party R&V Systems",to all affected customers. On September 28, 2007, Elekta issued a second Important Notice A304 which reiterated the same information as was in Important Notice A302. The notices described the product, problem and actions to be taken. This notice provides a workaround on how to avoid the issue. In order to avoid this issues documented, Elekta recommended that the customers set the tolerance for collimators in the iCom tolerance table to be 3.0 degrees-no less than 3.0 and to use the Linac Record, if configured, as a means of checking the final collimator position. Elekta noted that the issue will be resolved in a future release of Desktop Pro. If you any questions related to this issue, please contact your local Elekta, third party R&V system representative or Elekta Limited at +44 (0)1293 654200.
Quantity
229 units
Distribution
Worldwide distribution: USA including Puerto Rico and countries including: Canada and Mexico.
Lot, serial or model codes
134615, 134621, 134091, 136364, 134172, 136285, 134605, 134183, 136167, 136234, 134259, 134115, 134081, 136047, 136063, 134701, 134328, 134391, 134432, 134436, 134438, 136119, 134535, 134561, 134475, 134564, 134154, 134269, 134348, 134399, 136084, 136087, 136100, 134200, 134118, 134822, 134018, 136368, 134241, 134606, 134711, 134473, 134787, 134768, 134571, 134237, 134104, 136009, 134331, 134272, 136322, 136175, 136200, 134503, 136257, 134322, 134781, 134380, 134709, 136108, 134250, 134254, 134258, 134206, 134195, 134094, 134652, 134704, 136107, 134469, 134472, 134483, 134519, 134729, 136109, 134236, 134239, 134197, 134456, 134047, 134092, 134211, 136373, 134275, 134679, 136106, 136163, 134667, 136181, 134169, 134329, 134126, 134090, 134358, 134427, 134626, 134819, 134027, 134265, 134597, 136319, 134220, 134256, 134396, 134804, 134235, 136338, 134137, 134234, 134249, 136034, 136065, 136110, 136267, 136291, 134213, 134219, 136330, 134750, 136097, 136270, 134457, 134078, 134129, 136195, 134375, 134603, 136074, 134075, 134565, 134196, 134428, 134430, 134008, 134382, 134070, 134157, 134772, 136080, 134860, 134401, 134368, 134388, 134631, 134318, 136015, 136096, 134563, 136165, 136240, 134441, 134017, 134048, 134189, 134395, 134324, 134595, 134140, 134212, 136086, 134354, 136178, 136317, 136318, 134445, 136012, 136332, 134321, 136071, 136089, 136292, 136293, 134673, 134685, 136059, 136169, 134073, 134748, 134754, 134378, 134383, 134458, 134192, 134499, 136069, 136092, 134642, 134295, 134671, 136046, 136147, 134062, 136049, 136036, 136152, 136157, 136083, 136081, 134770, 136118, 136156, 136172, 134352, 134548, 134533, 136153, 136269, 136197, 136252, 136286, 136315, 136241, 134811, 134812, 136258, 136312, 134069, 136310, 134864, 134040, 134046, 134307
510(k) numbers
["K982713"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register