Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the cardiac troponin I method on the Stratus CS analyzer.

Recall number
Z-1555-2009
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Siemens Stratus CS Acute Care Troponin I CalPak. In vitro product intended to be used for calibration of the cardiac troponin I method on the Stratus CS analyzer.
Status
Terminated
Initiated
2008-04-28
Posted
not on file
Terminated
2009-07-14
Reason
The firm received and confirmed that some lots of Cardiac Troponin I (cTnl) CalPaks (CCTNI-CR) may exhibit calibration issues that cause lower than typical signal during calibration with the current population of cTnl TestPaks (CCTNI). This issue may result in failed calibrations due to "Slope Error" messages or falsely elevated patient sample and quality control (QC) results on calibrations that
Root cause
Labeling False and Misleading
Action
Siemens Healthcare Diagnostics, Inc. issued an Urgent Field Safety Notice dated April 29, 2008 to all affected customers instructing them to discontinue use of 3 CalPak lots and reduce the shelf-life for the remaining lots. For further information, contact Siemens Healthcare Diagnostics, Inc. at 1-302-631-7672.
Quantity
4415 cartons
Distribution
Worldwide Distribution - including United States, United Arab Emirates, Austria, Bermuda, Belize, Canada, Germany, Egypt, India, Japan, Saudi Arabia and Bengal.
Lot, serial or model codes
Lot Numbers: 637218002 (expired immediately), 637281002 (expired immediately), 637350002 (expired immediately), 638035002 (expired June 4, 2008); and 638105002 (expired August 14, 2008).
510(k) numbers
["K051650"]
PMA numbers
not on file
Product code
MMI
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register