Philips Medical Systems, Inc.: Philips Healthcare DigitalDiagnost System X-Ray

Recall number
Z-1554-2015
Recalling firm
Philips Medical Systems, Inc.
Product
Philips Healthcare DigitalDiagnost System X-Ray
Status
Terminated
Initiated
2015-02-09
Posted
2015-05-19
Terminated
2016-06-14
Reason
The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
Root cause
Software design (manufacturing process)
Action
Philips Healthcare Planned action: 1. Philips healthcare will contact customers and initiate a software update to correct the software deficiency causing the problem. 2. Philips engineers will install the software. 3. Philips engineers will perform testing to ensure the software update was effective. 4. Corrections are to be made free of charge 5. Corrective actions should be completed by August 15, 2015 For further questions call (978) 659-7892
Quantity
54 DigitalDiagnost
Distribution
Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
Lot, serial or model codes
Digital Diagnost Stationary X-ray System Release 4.0.0, 4.0.1, 4.0.2
510(k) numbers
["K141736"]
PMA numbers
not on file
Product code
MQB
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register