Zimmer Inc.: Zimmer Dermacarriers II skin graft carriers, 6 to 1 dermacarrier, Rx, latex free, sterile, Qty 1, Zimmer, Dover, Ohio; REF 00-2195-014-00. Single use device for use during skin grafting procedures.

Recall number
Z-1552-2008
Recalling firm
Zimmer Inc.
Product
Zimmer Dermacarriers II skin graft carriers, 6 to 1 dermacarrier, Rx, latex free, sterile, Qty 1, Zimmer, Dover, Ohio; REF 00-2195-014-00. Single use device for use during skin grafting procedures.
Status
Terminated
Initiated
2008-04-03
Posted
2008-07-30
Terminated
2009-12-17
Reason
Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.
Root cause
Packaging process control
Action
Consignees were notified via letter (Urgent: Medical Device Correction) dated 4/3/08 to identify their stocks on hand, verify package sealing, to discard any non-sealed packages, mail the Business Reply Card regardless if there are any remaining affected products.
Quantity
99345 of all units
Distribution
Worldwide Distribution --- including USA and countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, Ecuador, England, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Largos, Lebanon, Lithuania, Malta, Mexico, Morocco, Netherlands, Nigeria, Panama, Peru, Poland, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunis, Turkey and Venezuela.
Lot, serial or model codes
All lots.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FZW
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register