Bard Peripheral Vascular Inc: PowerPort duo M.R.I. Implantable Port, with attachable 9.5F Polurethane Open-Ended Dual-Lumen Venous Catheter, REF 1829500, UDI: (01) 00801741027185 - Product Usage: The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

Recall number
Z-1550-2021
Recalling firm
Bard Peripheral Vascular Inc
Product
PowerPort duo M.R.I. Implantable Port, with attachable 9.5F Polurethane Open-Ended Dual-Lumen Venous Catheter, REF 1829500, UDI: (01) 00801741027185 - Product Usage: The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
Status
Terminated
Initiated
2021-03-25
Posted
not on file
Terminated
2023-04-10
Reason
Catheters may experience difficulty in flushing, infusion, and/or aspiration, and septum dislodgements.
Root cause
Process control
Action
On March 25,2021 Bard Peripheral Vascular distributed an "Urgent: Medical Device Recall notification to affected customers via FedEx and E-Mail. In addition to informing consignees about the recall, the firm ask customer to take the following actions: 1. Please check all inventory locations within your institution for the specific catalog and lot numbers of the PowerPort duo M.R.I. Implantable Ports listed above and immediately destroy all affected product remaining in your possession. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Urgent Medical Device Recall. 3. If you purchased this product from a distributor, contact your distributor for further instructions. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the recall notification, whether or not you have any affected product, so that BD may acknowledge your receipt of this notification and process any applicable replacement orders. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 6. For customer support contact: 1-844-823-5433, Say "Recall" when prompted, M-F 8:00AM - 5:00 PM CT Actions Taken by BD: 1. BD will process replacement orders for all customers affected by the recall following receipt of the completed Customer Response Form. 2. Based on inventory levels there may be a delay of 6 weeks or more in the fulfillment of replacement orders. BD understands that supply interruptions can impact our customers ability to provide the best care for their patients and takes this matter very seriously, as such we fulfill orders as quickly as possible. 3. BD has implemen
Quantity
2,194 units
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN MO, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV.
Lot, serial or model codes
Code: REEU3063, REEV0989, REEV2633, REEX1380, REEZ1991
510(k) numbers
["K090512"]
PMA numbers
not on file
Product code
LJT
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register