FDA Medical Device Recalls (openFDA)

Stryker Craniomaxillofacial Division: Inion 1.5 x 6mm CPS(R) Baby Screw, 1.5mm Inion CPS(R) Baby 5+E, REF SCR 1221, 5/pack, STERILE/R Stryker Craniomaxillofacial, Portage, MI. The INION CPS BABY 1.5 BIOABSORBABLE FIXATION SYSTEM is a craniomaxillofacial plating system intended to maintain accurate alignment of bone fractures and osteotomies in the cranium and mid-face.

Recall number
Z-1550-2010
Recalling firm
Stryker Craniomaxillofacial Division
Product
Inion 1.5 x 6mm CPS(R) Baby Screw, 1.5mm Inion CPS(R) Baby 5+E, REF SCR 1221, 5/pack, STERILE/R Stryker Craniomaxillofacial, Portage, MI. The INION CPS BABY 1.5 BIOABSORBABLE FIXATION SYSTEM is a craniomaxillofacial plating system intended to maintain accurate alignment of bone fractures and osteotomies in the cranium and mid-face.
Status
Terminated
Initiated
2010-03-05
Posted
2010-05-07
Terminated
2012-06-06
Reason
During a sampling inspection, 2 lots of screws did not meet shear and torsion testing specifications.
Root cause
Nonconforming Material/Component
Action
The firm, Stryker, issued two "URGENT PRODUCT RECALL NOTIFICATION" letters, dated March 5, 2010 and March 19, 2010, to customers. The letters describe the product, problem and action to be taken by the customers. The customers were instructed to examine their inventory for the catalogue and lot numbers, return the enclosed confirmation indicating the location of the affected product, return any remaining product to Stryker at the noted address, and provide a copy of this notification to any site that they may have distributed the product to. If you have any further questions, contact Rob Yamashita, Sr. Regulatory Affairs Rep at (269) 323-4258 or email: rob.yamashita@stryker.com
Quantity
184 packs (5/pack)
Distribution
Nationwide distribution: CA, CO, CT, FL, GA, IL, LA, MA, MD, ME, MO, MT, NC, NH, NJ, NM, NY, OH, PA, RI, SC, TX and WI.
Lot, serial or model codes
Lot numbers 0808009 and 0809069. Model number SCR1221.
510(k) numbers
["K010352","K020266"]
PMA numbers
not on file
Product code
HWC
Firm FEI
3005101424
Firm city
Portage
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register