FDA Medical Device Recalls (openFDA)
Stryker Craniomaxillofacial Division: Inion 1.5 x 6mm CPS(R) Baby Screw, 1.5mm Inion CPS(R) Baby 5+E, REF SCR 1221, 5/pack, STERILE/R Stryker Craniomaxillofacial, Portage, MI. The INION CPS BABY 1.5 BIOABSORBABLE FIXATION SYSTEM is a craniomaxillofacial plating system intended to maintain accurate alignment of bone fractures and osteotomies in the cranium and mid-face.
- Recall number
- Z-1550-2010
- Recalling firm
- Stryker Craniomaxillofacial Division
- Product
- Inion 1.5 x 6mm CPS(R) Baby Screw, 1.5mm Inion CPS(R) Baby 5+E, REF SCR 1221, 5/pack, STERILE/R Stryker Craniomaxillofacial, Portage, MI. The INION CPS BABY 1.5 BIOABSORBABLE FIXATION SYSTEM is a craniomaxillofacial plating system intended to maintain accurate alignment of bone fractures and osteotomies in the cranium and mid-face.
- Status
- Terminated
- Initiated
- 2010-03-05
- Posted
- 2010-05-07
- Terminated
- 2012-06-06
- Reason
- During a sampling inspection, 2 lots of screws did not meet shear and torsion testing specifications.
- Root cause
- Nonconforming Material/Component
- Action
- The firm, Stryker, issued two "URGENT PRODUCT RECALL NOTIFICATION" letters, dated March 5, 2010 and March 19, 2010, to customers. The letters describe the product, problem and action to be taken by the customers. The customers were instructed to examine their inventory for the catalogue and lot numbers, return the enclosed confirmation indicating the location of the affected product, return any remaining product to Stryker at the noted address, and provide a copy of this notification to any site that they may have distributed the product to. If you have any further questions, contact Rob Yamashita, Sr. Regulatory Affairs Rep at (269) 323-4258 or email: rob.yamashita@stryker.com
- Quantity
- 184 packs (5/pack)
- Distribution
- Nationwide distribution: CA, CO, CT, FL, GA, IL, LA, MA, MD, ME, MO, MT, NC, NH, NJ, NM, NY, OH, PA, RI, SC, TX and WI.
- Lot, serial or model codes
- Lot numbers 0808009 and 0809069. Model number SCR1221.
- 510(k) numbers
- ["K010352","K020266"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 3005101424
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28