Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corp.: Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.

Recall number
Z-1550-06
Recalling firm
Roche Diagnostics Corp.
Product
Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
Status
Terminated
Initiated
2006-09-15
Posted
2006-09-29
Terminated
2008-04-30
Reason
Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method.
Root cause
Other
Action
Consignees were informed via letter dated 9/15/06 to perform Folate assays in batch mode and to run controls and liquid flow cleaning upon completion of each batch run. Distributors were requested to notify their customers.
Quantity
5688 units
Distribution
Nationwide.
Lot, serial or model codes
Lots 175646 and 174558.
510(k) numbers
["K043318"]
PMA numbers
not on file
Product code
CGN
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register