Roche Diagnostics Corp.: Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
- Recall number
- Z-1550-06
- Recalling firm
- Roche Diagnostics Corp.
- Product
- Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
- Status
- Terminated
- Initiated
- 2006-09-15
- Posted
- 2006-09-29
- Terminated
- 2008-04-30
- Reason
- Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method.
- Root cause
- Other
- Action
- Consignees were informed via letter dated 9/15/06 to perform Folate assays in batch mode and to run controls and liquid flow cleaning upon completion of each batch run. Distributors were requested to notify their customers.
- Quantity
- 5688 units
- Distribution
- Nationwide.
- Lot, serial or model codes
- Lots 175646 and 174558.
- 510(k) numbers
- ["K043318"]
- PMA numbers
- not on file
- Product code
- CGN
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28