FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: MEDLINE ReNewal ENT Coblator II PROcise mAx,W/ Integrated Cable Suction & Saline (Blue), Item Number REF EICA88980R

Recall number
Z-1549-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
MEDLINE ReNewal ENT Coblator II PROcise mAx,W/ Integrated Cable Suction & Saline (Blue), Item Number REF EICA88980R
Status
Open, Classified
Initiated
2025-03-05
Posted
2025-04-09
Terminated
not on file
Reason
Medline ReNewal has identified that the drip chambers of affected devices may disconnect from the saline line tubing when under tension.
Root cause
Under Investigation by firm
Action
Medline issued a "Medline ReNewal IMMEDIATE ACTION REQUIRED" notice dated 3/5/2025, to its consignees on 03/05/2025 via USPS first class mail. The notice explained the issue, potential risk, and requested the following actions: REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers. Quarantine all affected product. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Upon receipt of your submitted response form, FedEx return label(s) will be provided. Even if you do not have any affected product, please complete and return the form, as indicated on the response form. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline ReNewal. For questions, please contact 866-359-1704 or recalls@medline.com.
Quantity
224 units
Distribution
US Nationwide distribution
Lot, serial or model codes
UDI/DI 10080196415106, Lot Numbers: 506549, 508766, 509420, 512036, 512983, 514060, 514709, 516475, 516802, 517626, 518478, 518961, 519248, 519259, 519411, 519412, 519413, 519966, 519969, 520277, 522144, 522145, 527110, 527112, 528454, 528458, 528584, 531021, 531022.
510(k) numbers
["K012669"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register