Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.

Recall number
Z-1548-2013
Recalling firm
Biomet, Inc.
Product
Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
Status
Terminated
Initiated
2013-05-15
Posted
2013-06-17
Terminated
2014-06-09
Reason
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.
Root cause
Device Design
Action
On 5/15/2013 an URGENT MEDICAL DEVICE RECALL notification was sent to all consignees detailing to immediately locate and discontinue use of the product and return to Biomet. The following actions are REQUIRED: -Immediately locate and remove the identified device(s) listed below from circulation . -Carefully follow the instructions on the enclosed "FAX Back Response Form" . -Fax a copy of the Response Form to 574-372-1683 prior to return of product. -Use priority carrier for your shipment. - If you have further distributed this product. you MUST notify hospital personnel of this action via the enclosed "Dear O.R. Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices.
Quantity
17
Distribution
Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.
Lot, serial or model codes
Catalog number: 010001814 and lot numbers: 318520 and 644720.
510(k) numbers
["K080642"]
PMA numbers
not on file
Product code
PAO
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register