Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: NM/CT 870 DR Nuclear Medicine / CT Scanners

Recall number
Z-1547-2021
Recalling firm
GE Healthcare, LLC
Product
NM/CT 870 DR Nuclear Medicine / CT Scanners
Status
Open, Classified
Initiated
2021-02-07
Posted
2021-05-11
Terminated
not on file
Reason
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Root cause
Device Design
Action
GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
242 in total (17 units US)
Distribution
Worldwide distribution - US Nationwide
Lot, serial or model codes
GTIN*00840682140836
510(k) numbers
["K093514"]
PMA numbers
not on file
Product code
KPS
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register