Skip to the content

FDA Medical Device Recalls (openFDA)

Brainlab AG: BrainLAB M3 Micro-Multileaf-Collimator for Elekta/Philips Type III with MLC, model 41600-06, -6 A; BrainLAB AG, Acetylesterase 12, 85622 Edicendi, Germany. The m MLC is intended to assist the radiation oncologist in the delivery of radiation to well-defined target volumes while sparing surrounding normal tissue and critical organs from access radiation.

Recall number
Z-1547-2009
Recalling firm
Brainlab AG
Product
BrainLAB M3 Micro-Multileaf-Collimator for Elekta/Philips Type III with MLC, model 41600-06, -6 A; BrainLAB AG, Acetylesterase 12, 85622 Edicendi, Germany. The m MLC is intended to assist the radiation oncologist in the delivery of radiation to well-defined target volumes while sparing surrounding normal tissue and critical organs from access radiation.
Status
Terminated
Initiated
2009-03-19
Posted
2009-07-02
Terminated
2010-08-19
Reason
The Integrated Multileaf-Collimator (MLC) of the linear accelerator could interfere with the intended conformal fields of the BrainLab m3 micro-Multileaf-Collimator (mMLC) during patient treatment.
Root cause
Other
Action
BrainLAB AG issued an "Urgent Field Safety Notice/Product Notification" letter dated March 19, 2009 informing all customers of the affected device. Consignees were instructed on proper steps to take to avoid further issues. In addition, the most current m3 mMLC user guide, version 7.4, was scheduled to be sent to users in April 2009. For further questions, contact BrainLAB AG by phone at 1-800-597-5911).
Quantity
1 unit
Distribution
Nationwide Distribution - AL, CA, CO, FL, GA, IL, KS, KY, LA, MD, MI, MN, MS, ME, NJ, NY, OR, TN, TX, UT, VA, WV, WA and WI.
Lot, serial or model codes
Model 41600-6A; Serial Number 41600-6A/24143.
510(k) numbers
["K004022"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3002619595
Firm city
Kirchheim B. Muenchen
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register