FDA Medical Device Recalls (openFDA)
Lorad, A Hologic, Inc.: Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
- Recall number
- Z-1546-2008
- Recalling firm
- Lorad, A Hologic, Inc.
- Product
- Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
- Status
- Terminated
- Initiated
- 2008-02-28
- Posted
- 2008-05-28
- Terminated
- 2009-02-03
- Reason
- Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.
- Root cause
- Software design
- Action
- Consignees were notified by an Urgent Recall Notice sent on 3/21/08. The letter informed users that representatives were dispatched to all locations and they have corrected the magnification factor problem. The letter instructed users to re-evaluate any cases read using CAD from the period of installation until the date of system correction. The letter also instructed users that only images read with CAD markers are affected. The letter requests users to return a confirmation of notification. For additional information, contact 1-877-371-4372.
- Quantity
- 73 units
- Distribution
- Worldwide Distribution including USA and countries of Canada and the Republic of Korea.
- Lot, serial or model codes
- Serial Numbers: 28401060607, 28401060617, 28401060619, 28401060641, 28401060645, 28401071345, 28401071361, 28401071363, 28401082760, 28402050317, 28402071431, 28403060709, 28403071548, 28403071601, 28404060755, 28405060792, 28405060813, 28405071807, 28405071813, 28406071833, 28407050431, 28407060899, 28407071951, 28407072014, 28408050465, 28408072100, 28408072106, 28408072130, 28409030075, 28409030076, 28409050497, 28410072360, 28411061182, 28411061271, 28411072479, 28412050577, 28412050589, 28412061309, 28412061319, 28412061333, 28412061336, 28412061339, 28412072602, 29408072073, 28407072007S, 28409030072RF and 28409030077RM.
- 510(k) numbers
- not on file
- PMA numbers
- ["P010025"]
- Product code
- MUE
- Firm FEI
- 1220984
- Firm city
- Danbury
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28