FDA Medical Device Recalls (openFDA)

Lorad, A Hologic, Inc.: Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810

Recall number
Z-1546-2008
Recalling firm
Lorad, A Hologic, Inc.
Product
Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
Status
Terminated
Initiated
2008-02-28
Posted
2008-05-28
Terminated
2009-02-03
Reason
Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.
Root cause
Software design
Action
Consignees were notified by an Urgent Recall Notice sent on 3/21/08. The letter informed users that representatives were dispatched to all locations and they have corrected the magnification factor problem. The letter instructed users to re-evaluate any cases read using CAD from the period of installation until the date of system correction. The letter also instructed users that only images read with CAD markers are affected. The letter requests users to return a confirmation of notification. For additional information, contact 1-877-371-4372.
Quantity
73 units
Distribution
Worldwide Distribution including USA and countries of Canada and the Republic of Korea.
Lot, serial or model codes
Serial Numbers: 28401060607, 28401060617, 28401060619, 28401060641, 28401060645, 28401071345, 28401071361, 28401071363, 28401082760, 28402050317, 28402071431, 28403060709, 28403071548, 28403071601, 28404060755, 28405060792, 28405060813, 28405071807, 28405071813, 28406071833, 28407050431, 28407060899, 28407071951, 28407072014, 28408050465, 28408072100, 28408072106, 28408072130, 28409030075, 28409030076, 28409050497, 28410072360, 28411061182, 28411061271, 28411072479, 28412050577, 28412050589, 28412061309, 28412061319, 28412061333, 28412061336, 28412061339, 28412072602, 29408072073, 28407072007S, 28409030072RF and 28409030077RM.
510(k) numbers
not on file
PMA numbers
["P010025"]
Product code
MUE
Firm FEI
1220984
Firm city
Danbury
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register