FDA Medical Device Recalls (openFDA)
CareFusion 303, Inc.: BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qualified and authorized user in a clinical settings.
- Recall number
- Z-1545-2025
- Recalling firm
- CareFusion 303, Inc.
- Product
- BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qualified and authorized user in a clinical settings.
- Status
- Open, Classified
- Initiated
- 2025-03-11
- Posted
- 2025-04-09
- Terminated
- not on file
- Reason
- Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.
- Root cause
- Software design
- Action
- On 3/11/2025, "URGENT MEDICAL DEVICE CORRECTION" notices were mailed and emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Lab Manager, Clinical Engineering, and other customers asking them to take the following actions: 1) Best practice is to verify the total dose being removed is accurate against the provider's order, and if available utilize barcode medication administration as an additional check to ensure the correct dose is retrieved prior to administration. 2) Customers can inventory the CUBIE pocket via a cycle count or blind count to correct the inventory when discrepancies occur. Customers can refer to the BD Pyxis MedBank Cabinet User Guide - 10000414987 Ver. 01, Version v2.4 for instructions. Refer to chapter six (6), page 103. 3)Customers should ensure backup procedures (which may include but are not limited to keys, emergency medication supplies, alternate devices in proximity, etc) are in place and staff have all been properly educated on them. 4) Share this letter with any facilities within your network that you may have distributed affected devices to. 5) Post this Recall Letter on or near affected products as a reminder of the recall and the steps/instructions that are needed to be followed. 6) Complete and return the Response Form via email to bdrc44@bd.com Firm is in the process of contacting all customers to schedule an appropriate time for the software upgrade to MedBank software version 1.7. If you require further assistance, contact the firm's technical support: 1-866-930-9251 Phone Hours: 24 hours, 7 days a week Website: https://bdx.my.site.com/CC360/s/impactedproducts?rn=MMS-24-5163/MMS-25-5183/5184
- Quantity
- 7357
- Distribution
- US Nationwide distribution: CA, TX, WI, CO, AK, VA, TN, WA, NC, NY, MS, FL, KS, MI, IN, MN, MT, KY, MO, UT, NJ, OH, PA, ND, OR, NE, SD, NV, OK, GA, MA, IA, HI, ME, AZ, IL, SC, LA, WV, AL, ID, NM, DC, NH, VT, CT, MD, AR, WY.
- Lot, serial or model codes
- REF/UDI-DI(GTIN)/Software Version: 169-143/10885403512568/2.4; 169-145/10885403512568/2.4; 169-144/10885403512568/2.4; 169-142/10885403512568/2.4; 139088-01/1.2.3.23; 155288-01/10885403508721/2.4; 169-137/10885403512544/2.4; 155311-01/10885403510847/2.4; 169-151/10885403512544/1.2.3.23; 138921-01/2.4; 155159-01/10885403504655/2.4; 169-101/10885403512544/2.4; 139122-01/2.4; 155160-01/10885403504662/2.4; 169-102/10885403512544/2.4; 138912-01/2.4; 155161-01/10885403504679/2.4; 169-103/10885403512544/2.4; 138902-01/2.4; 169-104/10885403512544/2.4; 138903-01/2.4; 155163-01/10885403504686/2.4; 169-105/10885403512544/2.4; 138908-01/2.4; 169-106/10885403512544/2.4; 138906-01/2.4; 155165-01/10885403504693/2.4; 169-107/10885403512544/2.4; 169-152/10885403512544/1.2.3.23; 138907-01/2.4; 155166-01/10885403504570/2.4; 169-108/10885403512544/2.4; 139131-01/2.4; 155167-01/10885403504587/2.4; 169-109/10885403512544/2.4; 139091-01/2.4; 155168-01/10885403504594/2.4; 169-110/10885403512544/2.4; 138909-01/2.4; 155169-01/10885403504600/2.4; 169-111/10885403512544/2.4; 138910-01/2.4; 155310-01/10885403510830/2.4; 169-112/10885403512544/2.4; 169-150/10885403512544/1.2.3.23; 138915-01/2.4; 155171-01/10885403504617/2.4; 169-113/10885403512544/2.4; 138943-01/2.4; 155081-01/10885403503870/2.4; 169-23/10885403512568/2.4; 138955-01/2.4; 155082-01/10885403503887/2.4; 169-24/10885403512568/2.4; 138924-01/2.4; 155083-01/10885403503894/2.4; 169-25/10885403512568/2.4; 138960-01/2.4; 155084-01/10885403503900/2.4; 169-26/10885403512568/2.4; 138961-01/2.4; 155085-01/10885403503917/2.4; 169-27/10885403512568/2.4; 138926-01/2.4; 155086-01/10885403503726/2.4; 169-28/10885403512568/2.4; 138927-01/2.4; 155087-01/10885403503733/2.4; 169-29/10885403512568/2.4; 155293-01/10885403509926/2.4; 169-138/10885403512568/2.4; 155303-01/10885403510779/2.4; 139082-01/2.4; 155088-01/10885403503740/2.4; 169-30/10885403512568/2.4; 155305-01/10885403510793/2.4; 139056-01/2.4; 155089-01/10885403503757/2.4; 169-31/10885403512568/2.4; 155304-01/10885403510786/2.4; 139045-01/2.4; 155090-01/10885403503764/2.4; 169-32/10885403512568/2.4; 138932-01/2.4; 155091-01/10885403503771/2.4; 169-33/10885403512568/2.4; 139084-01/2.4; 155092-01/10885403503788/2.4; 169-34/10885403512568/2.4; 138963-01/2.4; 155093-01/10885403503795/2.4; 169-35/10885403512568/2.4; 155294-01/10885403509933/2.4; 169-139/10885403512568/2.4; 138938-01/2.4; 155094-01/10885403503801/2.4; 169-36/10885403512568/2.4; 139125-01/2.4; 155095-01/10885403503818/2.4; 169-37/10885403512568/2.4; 138919-01/2.4; 155096-01/10885403503696/2.4; 169-38/10885403512568/2.4; 139126-01/2.4; 155097-01/10885403504198/2.4; 169-39/10885403512568/2.4; 139127-01/2.4; 155098-01/10885403503702/2.4; 169-40/10885403512568/2.4; 139043-01/2.4; 155099-01/10885403504204/2.4; 169-41/10885403512568/2.4; 138933-01/2.4; 155100-01/10885403504211/2.4; 169-42/10885403512568/2.4; 155299-01/10885403510526/2.4; 169-140/10885403512568/1.2.3.23; 139142-01/2.4; 155101-01/10885403504228/2.4; 169-43/10885403512568/2.4; 155302-01/10885403510762/2.4; 138959-01/2.4; 155102-01/10885403504235/2.4; 169-44/10885403512568/2.4; 138954-01/2.4; 155103-01/10885403504242/2.4; 169-45/10885403512568/2.4; 138934-01/2.4; 155104-01/10885403504259/2.4; 169-46/10885403512568/2.4; 138950-01/2.4; 155105-01/10885403503719/2.4; 169-47/10885403512568/2.4; 138947-01/2.4; 155106-01/10885403504266/2.4; 169-48/10885403512568/2.4; 138945-01/2.4; 155107-01/10885403504273/2.4; 169-49/10885403512568/2.4; 138931-01/2.4; 155108-01/10885403504280/2.4; 169-50/10885403512568/2.4; 138929-01/2.4; 155109-01/10885403504327/2.4; 169-51/10885403512568/2.4; 138946-01/2.4; 155110-01/10885403504297/2.4; 169-52/10885403512568/2.4; 138940-01/2.4; 155111-01/10885403504303/2.4; 169-53/10885403512568/2.4; 138952-01/2.4; 155112-01/10885403504099/2.4; 169-54/10885403512568/2.4; 138917-01/2.4; 155113-01/10885403504105/2.4; 169-55/10885403512568/2.4; 138918-01/2.4; 155114-01/10885403504112/2.4; 169-56/10885403512568/2.4; 138941-01/2.4; 155115-01/10885403504129/2.4; 169-57/10885403512568/2.4; 138905-01/2.4; 155116-01/10885403504136/2.4; 169-58/10885403512568/2.4; 138916-01/2.4; 155117-01/10885403504143/2.4; 169-59/10885403512568/2.4; 139141-01/2.4; 155118-01/10885403504150/2.4; 169-60/10885403512568/2.4; 138914-01/2.4; 155119-01/10885403504334/2.4; 169-61/10885403512568/2.4; 139063-01/2.4; 155120-01/10885403504167/2.4; 169-62/10885403512568/2.4; 138953-01/2.4; 155121-01/10885403504174/2.4; 169-63/10885403512568/2.4; 138936-01/2.4; 155122-01/10885403504181/2.4; 169-64/10885403512568/2.4; 138962-01/2.4; 155123-01/10885403504082/2.4; 169-65/10885403512568/2.4; 138944-01/2.4; 155124-01/10885403504075/2.4; 169-66/10885403512568/2.4; 138957-01/2.4; 155125-01/10885403503658/2.4; 169-67/10885403512568/2.4; 138922-01/2.4; 155126-01/10885403503665/2.4; 169-68/10885403512568/2.4; 138930-01/2.4; 155127-01/10885403503672/2.4; 169-69/10885403512568/2.4; 139128-01/2.4; 155128-01/10885403503689/2.4; 169-70/10885403512568/2.4; 138958-01/2.4; 155129-01/10885403504341/2.4; 169-71/10885403512568/2.4; 138920-01/2.4; 155130-01/10885403504358/2.4; 169-72/10885403512568/2.4; 138937-01/2.4; 155131-01/10885403504365/2.4; 169-73/10885403512568/2.4; 139086-01/2.4; 155132-01/10885403504372/2.4; 169-74/10885403512568/2.4; 138942-01/2.4; 155133-01/10885403503542/2.4; 169-75/10885403512568/2.4; 138949-01/2.4; 155134-01/10885403503559/2.4; 169-76/10885403512568/2.4; 139062-01/2.4; 155135-01/10885403503566/2.4; 169-77/10885403512568/2.4; 138939-01/2.4; 155136-01/10885403503573/2.4; 169-78/10885403512568/2.4; 138948-01/2.4; 155137-01/10885403503597/2.4; 169-79/10885403512568/2.4; 138956-01/2.4; 155138-01/10885403503603/2.4; 169-80/10885403512568/2.4; 138923-01/2.4; 155139-01/10885403503610/2.4; 169-81/10885403512568/2.4; 139163-01/2.4. 1119-00/10885403512704/2.4; 1139-00/10885403512704/2.4. 139054-01/2.4; 139055-01/2.4; 139053-01/2.4; 139041-01/2.4; 139046-01/2.4; 139038-01/2.4; 139040-01/2.4; 139049-01/2.4; 139044-01/2.4; 139039-01/2.4; 139051-01/2.4; 139164-01/2.4; 139052-01/2.4; 139070-01/2.4; 139071-01/2.4; 139061-01/2.4; 139066-01/2.4; 139065-01/2.4; 139058-01/2.4; 139060-01/2.4; 139067-01/2.4; 139064-01/2.4; 139059-01/2.4; 139068-01/2.4; 139165-01/2.4; 139069-01/2.4. 1144-00/2.4; 1137-00/2.4; 1147-00/10885403517181/2.4 1145-00/10885403517167/2.4
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BRY
- Firm FEI
- 2016493
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28