Physio-Control, Inc.: PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).

Recall number
Z-1544-2026
Recalling firm
Physio-Control, Inc.
Product
PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Infant/Child Reduced Energy Defibrillation Electrodes are pre gelled, self-adhesive, therapy electrodes that allow hands free defibrillation. These electrodes reduce the energy delivered to the patient by a factor of 4:1. The Infant/child reduced energy defibrillation electrodes are therapy electrodes designed to allow personnel who are trained on AED device operation and in basic life support, or other physician authorized emergency medical response systems, to safely and effectively defibrillate infants and children who are less than 8 years old or weigh less than 25kg (55lbs).
Status
Open, Classified
Initiated
2026-01-21
Posted
2026-03-10
Terminated
not on file
Reason
Due to pediatric defibrillator electrode delamination
Root cause
Unknown/Undetermined by firm
Action
On 01/21/2026, the firm sent via certified mail an "Urgent: Medical Device Recall" letter to customer informing them that Stryker was notified of complaints that the pediatric electrodes were experiencing gel delamination and could lead to a noisy or distorted ECG waveform and in addition may deliver energy to a patient without the electrode's gel barrier may result in burns. Customers are instructed to: 1. Immediately check internal inventory to locate the affected product as listed above. 2. Stop use and segregate and quarantine the affected product. 3. On the business reply form, indicate if an affected electrode is in their possession. 4. Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 5. Upon receipt of the completed business reply form, Stryker will contact you to arrange for replacement of the electrode. 6. A return material authorization (RMA) will be provided when a replacement electrode is sent to return the affected electrodes back to Stryker. 7. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until the correction has been completed. 8. Inform Stryker if any of the subject electrodes have been distributed to other organizations. If so, provide contact details, so Stryker can inform the recipients appropriately. For Questions - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
Quantity
16,499 electrodes (US/OUS: 8327/8172)
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Australia, Canada, Hong Kong, Japan, Mexico, Malaysia, Netherlands, New Zealand, and Taiwan.
Lot, serial or model codes
Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682483 Lot Numbers: 315303, 315639, 315858, 316007, 320529, 318931
510(k) numbers
not on file
PMA numbers
["P160012"]
Product code
MKJ
Firm FEI
3015876
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register