Agilent Technologies, Inc.: GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).

Recall number
Z-1544-2025
Recalling firm
Agilent Technologies, Inc.
Product
GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).
Status
Open, Classified
Initiated
2025-03-20
Posted
2025-04-08
Terminated
not on file
Reason
DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.
Root cause
Nonconforming Material/Component
Action
On 3/20/2025, "URGENT MEDICAL DEVICE RECALL" notices dated 3/20/2025 were emailed to customers who were asked to do the following: 1. Please check your inventory and identify if you have the affected product. 2. Confirm the number of affected kits in your possession and discard any unused kits. 3. Inform those who need to be aware of this notification within your organization or any other organization to which the affected product(s) have been transferred. Please ensure that your organization maintains awareness of this notice. 4. Complete and return the acknowledgement form via email to fieldactions@agilent.com This issue is limited only to chromosome 12, and the data generated can be analyzed using the CytoDx functionalities by displaying the moving average line in the Genome View, Chromosome View, or Gene View of the Triage View. If you have any questions, contact Agilent Technologies, Inc. via email: andrea.topper@agilent.com or call 408-386-7294.
Quantity
70
Distribution
US distribution to the state of: MA and OUS (International) distribution to countries of: Spain, Italy, Japan, Colombia, Poland
Lot, serial or model codes
UDI-DI: 05700571112601. REF/Lot: K1201-64105/0006798023, 5190-7317/0006793917
510(k) numbers
["K163367"]
PMA numbers
not on file
Product code
PFX
Firm FEI
2916205
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register