Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829

Recall number
Z-1544-2018
Recalling firm
Becton Dickinson & Co.
Product
BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829
Status
Terminated
Initiated
2018-03-02
Posted
not on file
Terminated
2018-07-02
Reason
The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.
Root cause
Error in labeling
Action
On March 2, 2018, the firm sent Urgent Medical Device Product Advisory notices to customers via UPS. The notice identified the errors and the correct information, as well as potential health risks. Customers were asked to take the following actions: 1) Share the information with all users of the product within their facility to ensure awareness; 2) Complete the customer response form and return to the recalling firm; and 3) Report any associated adverse health consequences to the recalling firm. The recalling firm is not asking customer to send back product or discard. The recalling firm will provide users with an updated insert when available. BD Customer/Technical Support: 800-638-8663, Monday-Friday, 8:30AM-5PM (EST) for US customers.
Quantity
13,898
Distribution
Distributed to accounts in 8 states: CA, FL, NJ, NY, OR, PA, VA, and WI. International distribution to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Europe, Hong Kong, Japan, Singapore, Taiwan, and Thailand.
Lot, serial or model codes
All since 08/20/2014
510(k) numbers
not on file
PMA numbers
not on file
Product code
PPM
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register