Skip to the content

FDA Medical Device Recalls (openFDA)

Materialise USA LLC: PATIENT SPECIFIC INSTRUMENTS N-K FLEX PIN GUIDES (FEMUR AND TIBIA) NONSTERILE ~ MUST BE STERILIZED PRIOR TO USE Patient ID ..- 0864-R3 Manufactured by Materialise, Plymouth, MI Distributed by Zimmer, Warsaw IN. Patient specific instrument is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient image scans.

Recall number
Z-1544-2012
Recalling firm
Materialise USA LLC
Product
PATIENT SPECIFIC INSTRUMENTS N-K FLEX PIN GUIDES (FEMUR AND TIBIA) NONSTERILE ~ MUST BE STERILIZED PRIOR TO USE Patient ID ..- 0864-R3 Manufactured by Materialise, Plymouth, MI Distributed by Zimmer, Warsaw IN. Patient specific instrument is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient image scans.
Status
Terminated
Initiated
2012-03-23
Posted
2012-05-14
Terminated
2013-01-23
Reason
March 23, 2012 a field representative reported patient specific --0242 L tibia guide case packaging contained a different patient specific ..-0864-R tibia guide. The two patient specific products had been switched in packaging. Neither products had been delivered to the end user.
Root cause
Labeling mix-ups
Action
Materialise telephoned both field representatives on March 23, 2012 and confirmed recalled product mix-up. Materialise requested both cases be returned and replacements were provided. No affected product was received by the end user.
Quantity
1 unit
Distribution
Nationwide Distribution including the states of Ohio and Minnesota.
Lot, serial or model codes
Lot 56529886, REF 005970-000-03 Exp 2012-09-08
510(k) numbers
["K111492"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3005718816
Firm city
Plymouth
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register