FDA Medical Device Recalls (openFDA)
Smiths Medical ASD Inc.: Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
- Recall number
- Z-1543-2024
- Recalling firm
- Smiths Medical ASD Inc.
- Product
- Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
- Status
- Open, Classified
- Initiated
- 2023-12-19
- Posted
- 2024-04-15
- Terminated
- not on file
- Reason
- Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
- Root cause
- Process control
- Action
- Smiths Medical issued an Updated UREGNT MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2023 via US mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: "Customer Required Actions - Medfusion Infusion Pumps: 1. Locate any affected Medfusion Syringe Pumps that may be in your possession by referring to the list of affected devices included with the Response Form. This list includes any specific pump model/serial number(s) your organization purchased that were manufactured or serviced with potentially affected Barrel Clamp Guides. Each pump has a unique serial number found on the label on the bottom of the pump. 2. You may continue to use the pumps but utilize the Syringe Verification Reference Tool originally included with the 2018 Notice (also provided as Attachment 2 of this communication) until pumps containing potentially affected Barrel Clamp Guides are repaired. Customer Required Actions - Individually Sold Barrel Clamp Guides: 1. Locate any affected Barrel Clamp Guide (Part Number G6000716) lots in your parts inventory or within the Medfusion Syringe Pump(s) you have repaired at your facility. NOTE - If you have purchased affected individually sold Barrel Clamp Guides, the Response Form included with this notice will indicate the order numbers shipped to your organization and a Return Label will be provided. 2. If any potentially affected Barrel Clamp Guides have already been installed in pumps, you may continue to use the pumps, but utilize the Syringe Verification Reference Tool originally included with the 2018 Notice (also provided as Attachment 2 of this communication) until the pumps installed with potentially affected Barrel Clamp Guides are repaired. If you have affected devices, please contact Smiths Medical for repair of the pump." "DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Ask them to c
- Quantity
- 1842 guide barrel clamps
- Distribution
- Worldwide distribution.
- Lot, serial or model codes
- Lot Numbers: P0407365, P0486670, P0561740
- 510(k) numbers
- ["K111386"]
- PMA numbers
- not on file
- Product code
- MRZ
- Firm FEI
- 3012307300
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28