FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716

Recall number
Z-1543-2024
Recalling firm
Smiths Medical ASD Inc.
Product
Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
Status
Open, Classified
Initiated
2023-12-19
Posted
2024-04-15
Terminated
not on file
Reason
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Root cause
Process control
Action
Smiths Medical issued an Updated UREGNT MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2023 via US mail and email. The notice explained the issue, potential risk, and requested the following actions be taken: "Customer Required Actions - Medfusion Infusion Pumps: 1. Locate any affected Medfusion Syringe Pumps that may be in your possession by referring to the list of affected devices included with the Response Form. This list includes any specific pump model/serial number(s) your organization purchased that were manufactured or serviced with potentially affected Barrel Clamp Guides. Each pump has a unique serial number found on the label on the bottom of the pump. 2. You may continue to use the pumps but utilize the Syringe Verification Reference Tool originally included with the 2018 Notice (also provided as Attachment 2 of this communication) until pumps containing potentially affected Barrel Clamp Guides are repaired. Customer Required Actions - Individually Sold Barrel Clamp Guides: 1. Locate any affected Barrel Clamp Guide (Part Number G6000716) lots in your parts inventory or within the Medfusion Syringe Pump(s) you have repaired at your facility. NOTE - If you have purchased affected individually sold Barrel Clamp Guides, the Response Form included with this notice will indicate the order numbers shipped to your organization and a Return Label will be provided. 2. If any potentially affected Barrel Clamp Guides have already been installed in pumps, you may continue to use the pumps, but utilize the Syringe Verification Reference Tool originally included with the 2018 Notice (also provided as Attachment 2 of this communication) until the pumps installed with potentially affected Barrel Clamp Guides are repaired. If you have affected devices, please contact Smiths Medical for repair of the pump." "DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Ask them to c
Quantity
1842 guide barrel clamps
Distribution
Worldwide distribution.
Lot, serial or model codes
Lot Numbers: P0407365, P0486670, P0561740
510(k) numbers
["K111386"]
PMA numbers
not on file
Product code
MRZ
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register