PerkinElmer Life and Analytical Sciences, Wallac, OY: Specimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current software version) Specimen Gate Screening Center is used for data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.

Recall number
Z-1541-2018
Recalling firm
PerkinElmer Life and Analytical Sciences, Wallac, OY
Product
Specimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current software version) Specimen Gate Screening Center is used for data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.
Status
Terminated
Initiated
2017-11-17
Posted
not on file
Terminated
2021-08-12
Reason
Potential errors in patient results generated by the Screening Center product that include both false negative and false positive results.
Root cause
Software design
Action
PerkinElmer sent an Urgent Medical Device Correction letter dated November 14,2017. Instructions: 1. Implement additional control measures concerning demographic information, by taking one of the following actions a. Fully enter demographics before loading specimens on instruments (i.e. specimens can be punched, but the assay plates cannot be loaded to the measurement instruments). b. If the above mentioned is not possible or if previously entered demographic information changes after loading the assay plates on the measurement instruments, recalculate all your results in Specimen Gate laboratory Result Viewer prior to assay acceptance. 2. Follow unreported specimens with available tools to avoid any delay in reporting screening results. For further questions, please call (317) 418-1735
Quantity
8
Distribution
Worldwide Distribution - US Distribution to the state of Georgia., and to the countries of : Belgium, Canada, Denmark, Italy, and United Kingdom.
Lot, serial or model codes
All software versions from 1.0 to 1.8 (current software version)
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
3002808498
Firm city
Turku
Firm state
not on file
Firm country
Finland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register