FDA Medical Device Recalls (openFDA)
CaridianBCT, Inc.: Trima Accel Automated Blood Collection System, catalog number 917000000, Caridian BCT, Lakewood, CO 80215 USA. Product Usage: Automated Blood Collection System for use in humans.
- Recall number
- Z-1541-2012
- Recalling firm
- CaridianBCT, Inc.
- Product
- Trima Accel Automated Blood Collection System, catalog number 917000000, Caridian BCT, Lakewood, CO 80215 USA. Product Usage: Automated Blood Collection System for use in humans.
- Status
- Terminated
- Initiated
- 2010-06-01
- Posted
- 2012-05-11
- Terminated
- 2012-07-25
- Reason
- Labeling clarification and software updates to include fixes for end of run summaries, increasing boot reliability, corrections to the double red blood cell count collection TBV limit, and to be in compliance with the most current FDA guidance for post donation platelet count targets.
- Root cause
- Software design
- Action
- CaridianBCT sent a Letter notification dated March 2011 to consignees The letter identified the upcoming software updates to version 6.0.1 as well as operations manual changes. For questions contact your CaridianBCT Representative, the CaridianBCT Support Center at +1-877-339-4228 or 1-303-231-4357 or your local CaridianBCT Customer Service office.
- Quantity
- 2278 units
- Distribution
- USA Nationwide Distribution
- Lot, serial or model codes
- All serial numbers and lots manufactured from 1997 to present.
- 510(k) numbers
- ["K090003"]
- PMA numbers
- not on file
- Product code
- KZE
- Firm FEI
- 1000117351
- Firm city
- Lakewood
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28