Philips Medical Systems: CareVue Chart Release C ,Versions C.00.00 through C.00.06 Product No. 862246. This device is a clinical information system intended for use in data collection, storage, and management with independent bedside devices, and other systems that are connected either directly or through networks. It is used when there is a need for generating patient records and computation of drug dosage.
- Recall number
- Z-1540-2008
- Recalling firm
- Philips Medical Systems
- Product
- CareVue Chart Release C ,Versions C.00.00 through C.00.06 Product No. 862246. This device is a clinical information system intended for use in data collection, storage, and management with independent bedside devices, and other systems that are connected either directly or through networks. It is used when there is a need for generating patient records and computation of drug dosage.
- Status
- Terminated
- Initiated
- 2008-03-04
- Posted
- 2008-08-23
- Terminated
- 2016-08-15
- Reason
- Medications prescribed for one patient were printed on the record of another patient.
- Root cause
- Software design
- Action
- Philips Medical issued letter notification, dated 29 Feb 2008 to customers titled; "Urgent Medical Device Correction Notice". Customers were asked to not solely rely on the Worklist to remind them when to administer orders or interventions until the software has been upgraded. Customers should review the reminders and verify orders and interventions on the Medication Administration Record and Administration record. The correction will be in the form of a software upgrade and will be performed remotely by Philips service personnel.
- Quantity
- 138 for products
- Distribution
- Nationwide Canada, Australia, Austria, Belgium, Denmark., France, Finland, Germany, Great Britain, Italy, Malaysia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, and UAE
- Lot, serial or model codes
- Softwarre versions: C.00.00 through C.00.06
- 510(k) numbers
- ["K992636"]
- PMA numbers
- not on file
- Product code
- DXJ
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28