Skip to the content

FDA Medical Device Recalls (openFDA)

Masimo Corporation: Masimo Rad-G, Pulse Oximeter (W/Patient Cable), REF:9895 Rx Only

Recall number
Z-1539-2024
Recalling firm
Masimo Corporation
Product
Masimo Rad-G, Pulse Oximeter (W/Patient Cable), REF:9895 Rx Only,
Status
Open, Classified
Initiated
2024-02-15
Posted
2024-04-12
Terminated
not on file
Reason
Their is a potential that Oximeter may automatically power off and on resulting in loss of monitoring.
Root cause
Material/Component Contamination
Action
On February 14, 2024 Masimo issued a "Urgent: Medical Device Recall Notice" via FedEx and US Mail to affected consignees. On February 29, 2024, Masimo issued an updated communication to provide additional information. Masimo's asked consignees to take the following actions: " Please assess your inventory to confirm if you have Rad-G devices listed in Attachment 2-part number(s) and serial number(s). " If you identify affected Rad-G devices, please promptly remove them from use and follow the steps below. " Contact Masimo Technical Services to request return of the affected part and serial number (s) for repair or replacement. " Masimo s Technical Services is available Monday through Friday 6:00am -5:00pm PST for live assistance or 24 hours a day to leave a message and someone will contact you. Local contact information can be found at http://service.masimo.com. Complete Attachment 1, then email it to CustomerNotice@masimo.com, or Fax it to 1-949-297-7575 by MM DD, 2023. Starting on 8/13/24, additional recall notices were sent identifying additional affected devices. Distributors were asked to identify their direct customers and provide them with the recall notice.
Quantity
65
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Belize, Bolivia, Burkina Faso, Burundi, Cambodia, Canada, Chile, Colombia, Congo, The Democratic Republic of, Costa Rica, Cyprus, Czech Republic, Ethiopia, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Israel, Italy, Japan, Lesotho, Mozambique, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Poland, Puerto Rico, Qatar, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Tanzania, United Republic Of, Thailand, Trinidad And Tobago, Uganda, United Kingdom, Uruguay, Yemen, Zimbabwe.
Lot, serial or model codes
UDI:(01)00843997013789/ Serial Numbers: 6000083849 6000083870 6000083875 6000083898 6000083909 6000083932 6000083941 6000083959 6000083974 6000083978 6000083983 6000083991 6000083997 6000083999 6000084011 6000084018 6000084020 6000084042 6000084055 6000084075 6000084084 6000084090 6000084101 6000084106 6000084112 6000084132 6000084133 6000084157 6000084163 6000084181 6000084205 6000084211 6000084222 6000084236 6000084276 6000084290 6000084364 6000084392 6000084394 6000084429 6000084439 6000084442 6000084454 6000084455 6000084457 6000084513 6000084536 6000084543 6000084567 6000084574 6000084575 6000084607 6000084613 6000084647 6000090319 6000090325 6000090331 6000090333 6000090339 6000084465 6000083931 6000083954 6000083963 6000084250 6000084309
510(k) numbers
["K213676"]
PMA numbers
not on file
Product code
DQA
Firm FEI
3011353843
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register