Siemens Healthcare Diagnostics, Inc.: ADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).

Recall number
Z-1539-2020
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
ADVIA Chemistry Total Bilirubin_2 assay- SMN: 10341115 (40 mL) & 10341113 (70 mL). Product Usage: ADVIA TBIL_2 is for in vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma of adults and neonates on the ADVIA Chemistry systems. Measurement of total bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gallbladder block. A total bilirubin measurement in newborn infants is intended to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
Status
Terminated
Initiated
2020-02-13
Posted
not on file
Terminated
2021-04-23
Reason
Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy
Root cause
Under Investigation by firm
Action
Siemens Healthineers issued Urgent Medical Device Correction (UMDC # ACHC-20-02.A.US, ACHC-20-02.A.US.AC, ACHC-20-02.A.US.DM, ACHC-20-02.A.US.DV) and Urgent Field Safety Notice (UFSN # ACHC-20-02.A.OUS, ACHC-20-02.A.OUS.AC, ACHC-20-02.A.OUS.DM, ACHC-20-02.A.OUS.DV) on February 13, 2020 to all Atellica CH, ADVIA CH, Dimension, and Dimension Vista customers to inform them of the issue, the potential risk to health, and actions to be taken by the customer. Customers were instructed to:Review the information on eltrombopag interference (ADVIA CH and Atellica CH bilirubins) or use of total bilirubin assay is not recommended for patients on eltrombopag therapy (Dimension and Dimension Vista). Review the letter with their Medical Director Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. If they have received any complaints of illness or adverse events associated with the product, immediately contact their local Siemens Customer Care Center or their local Siemens Technical Support Representative.
Quantity
5666 US and 52202 OUS
Distribution
Nationwide Foreign: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom, Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Azerbaijan, Bahamas, Bahrain, Bangladesh, Benin, Bosnia Herzegovina, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Ecuador, Egypt, French Guinea, Georgia, Guadeloupe, Hong Kong, India, Indonesia, Iraq, Iran, Israel, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Macedonia, Malawi, Malaysia, Mali, Mexico, Montenegro, Morocco, Mozambique, Myanmar, New Zealand, Nigeria, Oman, Pakistan, Peru, Philippines, Qatar, Republic if Yemen, Republic Korea, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, South Africa, Swaziland, Taiwan, Tanzania, Thailand, Tunisia, Turkmenistan, United Arab Emirates, Ukraine, Uruguay, Venezuela, Vietnam, Zimbabwe
Lot, serial or model codes
ADVIA CH TBIL_2 (40 mL) 492094 (01)00630414513072(10)492094(17)20210828  ADVIA CH TBIL_2 (40 mL) 486525 (01)00630414513072(10)486525(17)20210728  ADVIA CH TBIL_2 (40 mL) 478149 (01)00630414513072(10)478149(17)20210428  ADVIA CH TBIL_2 (40 mL) 472677 (01)00630414513072(10)472677(17)20210328  ADVIA CH TBIL_2 (40 mL) 469083 (01)00630414513072(10)469083(17)20210128  ADVIA CH TBIL_2 (40 mL) 462464 (01)00630414513072(10)462464(17)20201128  ADVIA CH TBIL_2 (40 mL) 455486 (01)00630414513072(10)455486(17)20200828  ADVIA CH TBIL_2 (40 mL) 452851 (01)00630414513072(10)452851(17)20200728  ADVIA CH TBIL_2 (40 mL) 447317 (01)00630414513072(10)447317(17)20200628  ADVIA CH TBIL_2 (40 mL) 436498 (01)00630414513072(10)436498(17)20200328  ADVIA CH TBIL_2 (70 mL) 492102 (01)00630414513058(10)492102(17)20210828  ADVIA CH TBIL_2 (70 mL) 492111 (01)00630414513058(10)492111(17)20210828  ADVIA CH TBIL_2 (70 mL) 486533 (01)00630414513058(10)486533(17)20210728  ADVIA CH TBIL_2 (70 mL) 478154 (01)00630414513058(10)478154(17)20210428  ADVIA CH TBIL_2 (70 mL) 478159 (01)00630414513058(10)478159(17)20210428  ADVIA CH TBIL_2 (70 mL) 493710 (01)00630414513058(10)493710(17)20210428  ADVIA CH TBIL_2 (70 mL) 490263 (01)00630414513058(10)490263(17)20210328  ADVIA CH TBIL_2 (70 mL) 490264 (01)00630414513058(10)490264(17)20210328  ADVIA CH TBIL_2 (70 mL) 469088 (01)00630414513058(10)469088(17)20210128  ADVIA CH TBIL_2 (70 mL) 469094 (01)00630414513058(10)469094(17)20210128  ADVIA CH TBIL_2 (70 mL) 462470 (01)00630414513058(10)462470(17)20201128  ADVIA CH TBIL_2 (70 mL) 462477 (01)00630414513058(10)462477(17)20201128  ADVIA CH TBIL_2 (70 mL) 455496 (01)00630414513058(10)455496(17)20200828  ADVIA CH TBIL_2 (70 mL) 455501 (01)00630414513058(10)455501(17)20200828  ADVIA CH TBIL_2 (70 mL) 452858 (01)00630414513058(10)452858(17)20200728  ADVIA CH TBIL_2 (70 mL) 452863 (01)00630414513058(10)452863(17)20200728  ADVIA CH TBIL_2 (70 mL) 452870 (01)00630414513058(10)452870(17)20200728  ADVIA CH TBIL_2 (70 mL) 447325 (01)00630414513058(10)447325(17)20200628  ADVIA CH TBIL_2 (70 mL) 447332 (01)00630414513058(10)447332(17)20200628  ADVIA CH TBIL_2 (70 mL) 436510 (01)00630414513058(10)436510(17)20200328  ADVIA CH TBIL_2 (70 mL) 436515 (01)00630414513058(10)436515(17)20200328
510(k) numbers
["K170065"]
PMA numbers
not on file
Product code
JFM
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register