Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Panta Nail, Rx only, Sterile

Recall number
Z-1539-2017
Recalling firm
Integra LifeSciences Corp.
Product
Panta Nail, Rx only, Sterile,
Status
Terminated
Initiated
2017-03-03
Posted
not on file
Terminated
2018-05-21
Reason
Voluntary recall/Removal of Panta and Panta XI Nails because there may be the potential that the internal and external package seal does not provide the level of sterility intended.
Root cause
Packaging change control
Action
Integra sent "Urgent Medical Device Recall Letters and Return Response Forms" dated March 3, 2017 to their distributors and customers. All customers were asked to examine their inventory to determine if they had the recalled lots and if they did they are instructed to stop using them immediately. Distributors are asked if the recalled product was further distributed, those customers should be notified and instructed to follow the recalling procedures. The return response form should be completed, even if they do not have any recalled lots, annotating the amount of product and list a record of the lot numbers. The return response form should be sent back to Integra via email to: FCA2@integralife.com or via fax to: 1-609-750-4220. Upon receipt by Integra, if your company has identified recalled lots, Customer Service will contact the customer and provide an RMA number and directions to return the product, as well as input an order to replace the quantity indicated on the return response form. Should any customer have any questions regarding the recall instructions, they can contact Customer Service at 1-800-654-2873. To return units under field inventory, customer can contact 1-888-601-0203, option 2. For further questions, please call (717) 309-3966.
Quantity
4,449 units
Distribution
US Nationwide and Internationally
Lot, serial or model codes
Panta Nails -  Reference/Model #500050ND - Lot #'s: F33A, F33B, F3BN, F3Y5, F3Y6, F4S4, F56Y, F5LZ, F64W, F7C8, F7C9, F8SZ, F9PC, FCUG, FCUH, FEDW, FEVR, FGAX, FGLN and FHLP;  Reference/Model #500080ND - Lot #'s: F0F1, F64X, F81Y, F81Z, FEVS, FGLP and FGSK;  Reference/Model #500150ND - Lot #'s: F15L, F15L/1, F1ZQ, F27U, F27V, F3Y7, F5M0,F60W, F64Y, F64Z, F6L3, F7CA, F7CB, F7CC, F7CD, F821, F822, F8T1, F8T2, F9D6, F9D7, FEDX, FFB9, FG4T, FGSL and FHLR;  Reference/Model #500180ND - Lot #'s: F1ZS, F33C, F3M9, F4H4, F4ZZ, F68B, F823, F824, F8T3, FDEC, FEDY, FFBA and FFQL;  Reference/Model #500250ND - Lot #'s: EP14, F0WQ, F15M, F4H5, F500, F5M1, F650, F825, F826, F827, F828, F8T4, FGLQ and FGSM;  Reference/Model #500280ND - Lot #'s: F27T, F33D, F3BP, F3MA, F4H6, F5M2, F6L4, F829, F82A, F82B, F82C, F8T5 and FFBB;  Reference/Model #500350ND - Lot #'s: E2W1/1, F3BQ, F3MB, F3MD, F3Y8, F4H7, F5M3, F651, F6TT and FF3B;  Reference/Model #500380ND - Lot #'s: F3BR, F3MC, F4S5, F5M4, F60X, F652, F82D, F82E, F8T6 and FAF2  Panta XL Nails -  Reference/Model #510111ND - Lot #'s: F3H8, F4XA, F507, F5M5, F68C, F82F, F82G, F8T7, F9DC and F9DD;  Reference/Model #51014ND - Lot #'s: ELSS, EPFZ, F3H9, F82H, F82J, F8LG, FEVT and FFBC;  Reference/Model #510211ND - Lot #'s: F6JA, F82K, F82L, F8C8, F8T8, FFQN and FG4V;  Reference/Model #510241ND - Lot #'s: F5M6, F82M, F82N, F8C9 and F8T9;  Reference/Model #510311ND - Lot #'s: EPG0/1, EPG0/2, EPG0/3, EPG0/4, EPG0/5, F82P, F82Q and F8TA;  Reference/Model #510341ND - Lot #'s: F4H8, F508, F657 and F8TB
510(k) numbers
["K050882","K091788"]
PMA numbers
not on file
Product code
HSB
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register