Heartware: HeartWare HVAD System Controller The HeartWare Ventricular Assist System (VAS is indicated for use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare Ventricular Assist System (VAS) is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter

Recall number
Z-1538-2017
Recalling firm
Heartware
Product
HeartWare HVAD System Controller The HeartWare Ventricular Assist System (VAS is indicated for use as a bridge-to-cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure. The HeartWare Ventricular Assist System (VAS) is designed for in-hospital and out-of-hospital settings, including transportation via fix-wing aircraft or helicopter
Status
Open, Classified
Initiated
2017-02-03
Posted
not on file
Terminated
not on file
Reason
Product recall of all HVAD Controllers currently on the market and to exchange them for the next generation HVAD Controller (Controller2.0)
Root cause
Software design
Action
The firm will release a communication to their consignees stating the following: As a reminder, as with all HVAD Controllers, continue to reinforce the following with your patients and staff at all opportunities: " Patients should continue to have a backup HVAD Controller ready at all times in the event of a primary HVAD Controller failure. " Staff only: The driveline extension cable is to be used during the pre-implant test only. It is not intended to be used after the pump is implanted in the patient. Hospital and Clinician Actions (to be executed in the following order): 1) Review the enclosed notice and forms, and forward the notice to those individuals within your organization who need to be aware of its contents. 2) Complete, sign, and return the Acknowledgement Form to HeartWare within thirty (30) days of receipt of this letter. 4) Quarantine and replace affected HVAD Controllers, DC Adapters, Instructions for Use, Emergency Responder Guides and Patient Manuals in hospital inventory after training is complete. For every patient, quarantine and replace the following under clinical supervision in an environment where appropriate support equipment is readily available: - Primary and Backup HVAD Controller; - Affected DC Adapters; and - Patient Manual and Emergency Responder Guide. Clinicians are reminded not to perform an HVAD Controller exchange during an active electrical fault alarm as the HVAD Pump will be running a single stator. If an electrical fault is present, download patient log files and contact your HeartWare representative to resolve the electrical fault before executing the controller exchange. 5) Return all quarantined HVAD Controllers and DC Adapters to HeartWare. Your HeartWare representative will assist you with this process. 6) Completion Form. Once affected product in inventory has been identified and returned, complete and return the attached Completion Form to CON2.0@heartware.com or your HeartWare representative no late
Quantity
8,343 US and 13,145 INTERNATIONAL
Distribution
Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Lithuania, , Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Turkey, United Arab Emirates, UK, and Vietnam.
Lot, serial or model codes
HeartWare HVAD System Controller Product numbers: 1401US, 1401XX, 1403US, 1407XX, and 1400 with the serial Number(s) lower than CON300000 HeartWare HVAD DC Adapter Product number: 1435  .
510(k) numbers
not on file
PMA numbers
["P100047"]
Product code
DSQ
Firm FEI
3007042319
Firm city
Miami Lakes
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register