Medtronic Inc. Cardiac Rhythm Disease Managment: Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN 55112. A microprocessor based, multichannel clot timing instrument with automated syringe handling for pipetting blood into single use cartridges. It performs in vitro heparin sensitivity evaluations, heparin assays, activated clotting times and platelet function evaluations.

Recall number
Z-1537-2010
Recalling firm
Medtronic Inc. Cardiac Rhythm Disease Managment
Product
Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN 55112. A microprocessor based, multichannel clot timing instrument with automated syringe handling for pipetting blood into single use cartridges. It performs in vitro heparin sensitivity evaluations, heparin assays, activated clotting times and platelet function evaluations.
Status
Terminated
Initiated
2010-03-08
Posted
2010-05-04
Terminated
2011-12-31
Reason
The heparin assay controls may run longer than normal and in some cases, fail to give an acceptable result.
Root cause
Process change control
Action
Medtronic issued an "Urgent Medical Device Notification" letter dated March 8, 2010 addressed to "Dear Valued Customer" and was sent to three locations at each customer site: Risk Management, Cardiovascular Surgery and Clinical Laboratory-Point of Care. The letter described the product and the issue, provided instructions for control and patient testing. The customer was requested to return an enclosed certificate as proof of notification. For further information, contact your Medtronic Sales Representative or call 1-763-391-9915.
Quantity
508 US, 36 OUS
Distribution
Worldwide Distribution -- United States, International West (countries included in Western Europe and the Middle East), Australia and Canada.
Lot, serial or model codes
HMS Plus Catalog Number: 30514;  Heparin Assay Controls: 306-01POR, 306-02POR, 306-03POR, 306-04POR, 306-05POR and 306-09POR.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JPA
Firm FEI
2182208
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register