Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neurological: Medtronic 3778 Lead Kit, 1x8 Compact Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis, MN 55432-5604.

Recall number
Z-1537-06
Recalling firm
Medtronic Neurological
Product
Medtronic 3778 Lead Kit, 1x8 Compact Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis, MN 55432-5604.
Status
Terminated
Initiated
2006-08-02
Posted
2006-09-29
Terminated
2006-12-31
Reason
Model 3777 Standard Octad Lead (Lot V009545) and Model 3778 Compact Octad Lead (Lot V009546) were mispackaged. The result of this error is that lead kits labeled 3777 lot V009545 may contain a 3778 compact-spaced lead, while those labeled 3778 lot V009546 may contain a 3777 standard-spaced lead. The result of this error is an overall increase or decrease of 14 mm in the electrode coverage.
Root cause
Other
Action
Medtronic representatives retrieved non-implanted devices. An Urgent Medical Device Recall letter was sent to physicians to describe the issue and patient management information.
Quantity
22 total
Distribution
Nationwide distribution --- including states of CA, GA, KY, LA, and TX.
Lot, serial or model codes
Model 3778, Lot V009546
510(k) numbers
not on file
PMA numbers
not on file
Product code
LGW
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28