Tornier S.A.S.: stryker Blueprint Mixed Reality Glenoid Digitizer, Catalog Number MRUE201, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthroplasty

Recall number
Z-1535-2025
Recalling firm
Tornier S.A.S.
Product
stryker Blueprint Mixed Reality Glenoid Digitizer, Catalog Number MRUE201, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthroplasty
Status
Open, Classified
Initiated
2025-03-05
Posted
2025-04-08
Terminated
not on file
Reason
Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its intended use. The most commonly reported complaints were related to impaired software functioning, impaired instrumentation functioning, and general difficulties using the product.
Root cause
Device Design
Action
Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 03/12/2025 via email. The notice explained the issue and requested the following: "1. Immediately check your internal inventory to locate the product listed on the attached business reply form, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Return the enclosed business reply form by email to confirm receipt of this notification/document product segregation. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers." For questions or concerns, please contact memphis.fieldaction@stryker.com. On 03/27/2025, Stryker issued an updated notice to include and additional product via email.
Quantity
22 units
Distribution
US: TX, MN, AK, PA, MA, TN, ID, and France, Canada, and Ireland
Lot, serial or model codes
All lots manufactured through Oct 2024, Distributed 23-Aug-2023 through 31-Dec-2024, GTIN 03700434022933
510(k) numbers
["K222510"]
PMA numbers
not on file
Product code
OLO
Firm FEI
3000931034
Firm city
Montbonnot-Saint-Martin
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register